Literature DB >> 22579442

Microbial contamination of non-invasive ventilation devices used by adults with cystic fibrosis.

A Mutagi1, E F Nash, S Cameron, G Abbott, P Agostini, J L Whitehouse, D Honeybourne, E Boxall.   

Abstract

BACKGROUND: There is currently little evidence regarding potential risks of bacterial contamination of non-invasive ventilation (NIV) devices used by cystic fibrosis (CF) patients. AIM: The aim of this study was to determine the extent of bacterial contamination of NIV devices in our regional adult CF centre.
METHODS: Seven NIV devices recently used by CF patients chronically infected with Pseudomonas aeruginosa or Burkholderia cepacia complex (BCC) were swabbed in seven areas, both external and internal. Two devices had undergone ethylene oxide (EtO) sterilization between patient use and swabbing, and five devices had not undergone EtO sterilization.
FINDINGS: Swabs from five devices had insignificant growth of environmental organisms and two devices had significant growth of environmental organisms. No CF pathogens were isolated from any machine.
CONCLUSIONS: No evidence was found of pathogenic microbial contamination of NIV devices used by CF patients in this small study. We suggest that further studies examine for evidence of bacterial contamination of NIV devices and that this issue should be included in future CF infection control guidelines.
Copyright © 2012 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22579442     DOI: 10.1016/j.jhin.2012.03.004

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  3 in total

1.  Exhaled gases and the potential for cross-infection via noninvasive ventilation machines.

Authors:  Benjamin Messer; Alison Dawn Armstrong; Nicholas David Lane; Ali Robb; Robert Edward Bullock
Journal:  ERJ Open Res       Date:  2022-07-04

2.  Persistent Pseudomonas aeruginosa infection associated with non-invasive ventilation in a child with spinal muscular atrophy type 1.

Authors:  Archana Chacko; Peter D Sly; Tim Kidd; Leanne Gauld
Journal:  J Clin Sleep Med       Date:  2021-03-01       Impact factor: 4.062

3.  Cleaning and infection control of airway clearance devices used by CF patients.

Authors:  Eynav Manor; Michal Gur; Yuval Geffen; Lea Bentur
Journal:  Chron Respir Dis       Date:  2017-05-17       Impact factor: 2.444

  3 in total