Literature DB >> 22573638

Clinical assessment of noninvasive intracranial pressure absolute value measurement method.

A Ragauskas1, V Matijosaitis, R Zakelis, K Petrikonis, D Rastenyte, I Piper, G Daubaris.   

Abstract

OBJECTIVE: To assess prospectively the accuracy and precision of a method for noninvasive intracranial pressure (ICP) measurement compared with invasive gold standard CSF pressure measurement.
METHODS: Included were 62 neurologic patients (37 idiopathic intracranial hypertension, 20 multiple sclerosis, 1 Guillain-Barré syndrome, 1 polyneuropathy, and 3 hydrocephalus). The average age was 40 ± 12 years. All patients had lumbar puncture indicated as a diagnostic procedure. ICP was measured using a noninvasive ICP measurement method, which is based on a two-depth high-resolution transcranial Doppler insonation of the ophthalmic artery (OA). The OA is being used as a natural pair of scales, in which the intracranial segment of the OA is compressed by ICP and the extracranial segment of the OA is compressed by extracranial pressure (Pe) applied to the orbit. The blood flow parameters in both OA segments are approximately the same in the scales balance case when Pe = ICP. All patients had simultaneous recording of noninvasive ICP values and invasive gold standard CSF pressure values.
RESULTS: Analysis of the 72 simultaneous paired recordings of noninvasive ICP and the gold standard CSF pressure showed good accuracy for the noninvasive method as indicated by the low mean systematic error (0.12 mm Hg; confidence level [CL] 0.98). The method also showed high precision as indicated by the low SD of the paired recordings (2.19 mm Hg; CL 0.98). The method does not need calibration.
CONCLUSION: The proposed noninvasive ICP measurement method is precise and accurate compared with gold standard CSF pressure measured via lumbar puncture.

Entities:  

Mesh:

Year:  2012        PMID: 22573638     DOI: 10.1212/WNL.0b013e3182574f50

Source DB:  PubMed          Journal:  Neurology        ISSN: 0028-3878            Impact factor:   9.910


  34 in total

1.  Measuring and monitoring ICP in Neurocritical Care: results from a national practice survey.

Authors:  DaiWai M Olson; Hunt H Batjer; Kamal Abdulkadir; Christiana E Hall
Journal:  Neurocrit Care       Date:  2014-02       Impact factor: 3.210

2.  Intracranial pressure changes after mild traumatic brain injury: a systematic review.

Authors:  Mohammad Nadir Haider; John J Leddy; Andrea L Hinds; Nell Aronoff; Diane Rein; David Poulsen; Barry S Willer
Journal:  Brain Inj       Date:  2018-04-27       Impact factor: 2.311

Review 3.  Non-invasive intracranial pressure assessment.

Authors:  Llewellyn C Padayachy
Journal:  Childs Nerv Syst       Date:  2016-07-21       Impact factor: 1.475

Review 4.  Non-invasive detection of intracranial pressure related to the optic nerve.

Authors:  Jian Li; Chao Wan
Journal:  Quant Imaging Med Surg       Date:  2021-06

5.  Non-invasive assessment of ICP in children: advances in ultrasound-based techniques.

Authors:  Llewellyn C Padayachy; C Robba; R Brekken
Journal:  Childs Nerv Syst       Date:  2019-10-04       Impact factor: 1.475

Review 6.  Noninvasive assessment of cerebrospinal fluid pressure.

Authors:  Beau B Bruce
Journal:  J Neuroophthalmol       Date:  2014-09       Impact factor: 3.042

7.  Can intracranial pressure be measured non-invasively bedside using a two-depth Doppler-technique?

Authors:  Lars-Owe D Koskinen; Jan Malm; Rolandas Zakelis; Laimonas Bartusis; Arminas Ragauskas; Anders Eklund
Journal:  J Clin Monit Comput       Date:  2016-03-14       Impact factor: 2.502

8.  Noninvasive intracranial pressure estimation by orbital subarachnoid space measurement: the Beijing Intracranial and Intraocular Pressure (iCOP) study.

Authors:  Xiaobin Xie; Xiaojun Zhang; Jidi Fu; Huaizhou Wang; Jost B Jonas; Xiaoxia Peng; Guohong Tian; Junfang Xian; Robert Ritch; Lei Li; Zefeng Kang; Shoukang Zhang; Diya Yang; Ningli Wang
Journal:  Crit Care       Date:  2013-07-24       Impact factor: 9.097

9.  From head micro-motions towards CSF dynamics and non-invasive intracranial pressure monitoring.

Authors:  Arnošt Mládek; Václav Gerla; Petr Šeba; Vladimír Kolář; Petr Skalický; Helen Whitley; Lenka Lhotská; Vladimír Beneš; Ondřej Bradáč
Journal:  Sci Rep       Date:  2021-07-12       Impact factor: 4.379

10.  The difference in translaminar pressure gradient and neuroretinal rim area in glaucoma and healthy subjects.

Authors:  Lina Siaudvytyte; Ingrida Januleviciene; Arminas Ragauskas; Laimonas Bartusis; Indre Meiliuniene; Brent Siesky; Alon Harris
Journal:  J Ophthalmol       Date:  2014-04-30       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.