| Literature DB >> 22567548 |
Ravi Kumar Konda1, B R Challa, Babu Rao Chandu, Kothapalli B Chandrasekhar.
Abstract
A simple, sensitive, and rapid HPLC-MS/MS method was developed and validated for quantitative estimation of memantine in human plasma. Chromatography was performed on Zorbax SB-C(18) (4.6 × 75 mm, 3.5 μm) column. Memantine (ME) and internal standard Memantine-d6(MED6) were extracted by using liquid-liquid extraction and analyzed by LC-ESI-MS/MS using multiple-reaction monitoring (MRM) mode. The assay exhibited a linear dynamic range of 50.00-50000.00 pg/ml for ME in human plasma. This method demonstrated an intra- and interday precision within the range of 2.1-3.7 and 1.4-7.8%, respectively. Further intra- and interday accuracy was within the range of 95.6-99.8 and 95.7-99.1% correspondingly. The mean recovery of ME and MED6 was 86.07 ± 6.87 and 80.31 ± 5.70%, respectively. The described method was successfully employed in bioequivalence study of ME in Indian male healthy human volunteers under fasting conditions.Entities:
Year: 2012 PMID: 22567548 PMCID: PMC3335321 DOI: 10.1155/2012/101249
Source DB: PubMed Journal: J Anal Methods Chem ISSN: 2090-8873 Impact factor: 2.193
Figure 1Chemical structures of (a) Memantine hydrochloride and (b) Memantine-D6HCL.
Figure 2(a) Mass spectra of Memantine parent ion (Q1). (b) Mass spectra of Memantine product ion (Q3). (c) Mass spectra of Memantine-D6 parent ion. (d) Mass spectra of Memantine-D6 product ion (Q3).
Figure 3(a) MRM chromatogram for blank plasma. (b) Chromatogram of LOQ.
Concentration data form validation.
| Spiked plasma concentration (pg/ml) | Concentration measured (pg/ml) Mean ± Sd | Precision (% CV) | Accuracy % |
|---|---|---|---|
| 50.00 | 49.92 ± 0.44 | 0.80 | 99.23 |
| 100.00 | 100.21 ± 1.91 | 1.90 | 98.14 |
| 500.00 | 502.73 ± 6.83 | 1.40 | 98.65 |
| 1000.00 | 1000.54 ± 11.53 | 1.10 | 98.92 |
| 5000.00 | 5005.06 ± 38.75 | 0.80 | 99.28 |
| 10000.00 | 9978.95 ± 160.56 | 1.60 | 98.45 |
| 20000.00 | 19871.04 ± 303.46 | 1.50 | 98.53 |
| 30000.00 | 29759.82 ± 508.47 | 1.70 | 98.37 |
| 40000.00 | 40310.88 ± 123.85 | 0.30 | 99.74 |
| 50000.00 | 50084.85 ± 266.72 | 0.50 | 99.55 |
Precision and accuracy (analysis with spiked plasma samples at three different concentrations).
| Spiked plasma concentration (pg/ml) | Within-run ( | Between-run ( | ||||
|---|---|---|---|---|---|---|
| Concentration measured ( | Precision (%CV) | Accuracy % | Concentration measured ( | Precision (%CV) | Accuracy % | |
| 150.00 | 143.40 ± 3.20 | 2.20 | 95.60 | 143.50 ± 3.90 | 2.70 | 95.70 |
| 15000.00 | 14746.40 ± 338.40 | 2.30 | 98.30 | 14719.10 ± 248.30 | 1.70 | 98.10 |
| 35000.00 | 34935.20 ± 730.40 | 2.10 | 99.80 | 34699.20 ± 498.40 | 1.40 | 99.10 |
Stability of Memantine in human plasma samples.
| Spiked plasma concentration (pg/ml | Concentration measured (pg/ml) | Precision (%CV) | Accuracy (%) |
|---|---|---|---|
| Room temperature stability for 26 hr in plasma | |||
| 150.00 | 151.23 | 1.00 | 100.82 |
| 35000.00 | 34926.50 | 0.80 | 99.79 |
| Three freeze-thaw cycles | |||
| 150.00 | 153.02 | 2.30 | 102.01 |
| 35000.00 | 34818.00 | 0.90 | 99.48 |
| Auto sampler stability for 79 hr | |||
| 150.00 | 154.59 | 1.40 | 103.06 |
| 35000.00 | 35213.50 | 0.90 | 100.61 |
| Stability for 76 days −30°C | |||
| 150.00 | 151.53 | 3.20 | 101.02 |
| 35000.00 | 35080.50 | 1.40 | 100.23 |
Figure 4Mean plasma concentrations of test versus reference after a 10 mg dose (one 10 mg tablet) single oral dose (20 healthy volunteers).
Mean pharmacokinetic parameters of Memantine in 20 healthy human volunteers after oral administration of 10 mg test and reference products.
| Pharmacokinetic details of Memantine in human plasma | ||
|---|---|---|
| Pharmacokinetic Parameter | Reference | Test |
| Mean ± SD | Mean ± SD | |
|
| 14368.57 ± 4044.16 | 14328 ± 4324.76 |
| AUC0− | 654545.5 ± 70423.12 | 674564.4 ± 67858.99 |
| AUC0− | 1053469.0 ± 77690.79 | 1136607 ± 74862.04 |
|
| 7.0 | 7.5 |
|
| 49.29 | 53.35 |
AUC0−: Area under the curve extrapolated to infinity.
AUC0−: Area under the curve up to the last sampling time.
C max: The maximum plasma concentration.
T max: The time to reach peak concentration.
Pharmacokinetic parameters of memantine after administration of 10 mg of test and reference products in 20 healthy human volunteers.
| Pharmacokinetic parameters |
| AUC0− | AUC0− |
|---|---|---|---|
| Test/Ref | 99.72 | 103.06 | 107.89 |