Arrhythmia discrimination using hemoglobin spectroscopy in humans.

S J Compton1, C D Swerdlow, R C Canby, G G Strobel, J D Zagrodzky, C Cinbis, J K Carney, S K Bhunia.   

Abstract

BACKGROUND: Inappropriate therapies are frequently delivered by implantable cardioverter-defibrillators (ICDs). We have investigated muscle perfusion as a means of augmenting arrhythmia discrimination by using implanted near-infrared spectroscopy.
OBJECTIVE: To evaluate hemodynamic stability by monitoring muscle perfusion from within the ICD pocket, in fresh tissue and inside the scar capsule on preexisting ICD generators, during induced cardiac arrhythmias, in humans.
METHODS: The sensor was implanted on or under the pectoral muscle, during ICD defibrillation threshold testing. A microvascular oxygenation trend indicator (O2 Index) was computed during 74 induced ventricular fibrillation and 34 normal sinus rhythm episodes in 34 patients and also during 28 atrial and 90 ventricular overdrive pacing episodes as simulations of supraventricular and ventricular tachycardias, respectively.
RESULTS: On average, the change in oxygenation, based on the O2 Index, was statistically significant (P <.003) from baseline within 3 seconds following cardiac arrest. An optimized O2 Index, used for detecting the hemodynamic trend, exhibited a decreasing trend during ventricular fibrillation (P <.0001) and was different from that during normal sinus rhythm (P <.0001). The sensitivity for the detection of ventricular fibrillation was 100%, and the specificity for the rejection of normal sinus rhythm was 82% in the presence of scar tissue on the optical sensor. For a 35-mm Hg drop in the mean arterial pressure as the threshold for hemodynamic instability, the specificity for the rejection of hemodynamically stable atrial and ventricular pacing episodes was 93% and 71%, respectively.
CONCLUSION: An implantable near-infrared spectroscopic sensor may be useful for hemodynamic monitoring during cardiac arrhythmias to prevent inappropriate therapy.
Copyright © 2012 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22561598     DOI: 10.1016/j.hrthm.2012.05.003

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  2 in total

1.  Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks.

Authors:  Daniel Keene; Matthew J Shun-Shin; Ahran D Arnold; James P Howard; David Lefroy; D Wyn Davies; Phang Boon Lim; Fu Siong Ng; Michael Koa-Wing; Norman A Qureshi; Nick W F Linton; Jaymin S Shah; Nicholas S Peters; Prapa Kanagaratnam; Darrel P Francis; Zachary I Whinnett
Journal:  JACC Clin Electrophysiol       Date:  2019-03-27

2.  Optimizing atrio-ventricular delay in pacemakers using potentially implantable physiological biomarkers.

Authors:  Daniel Keene; Alejandra A Miyazawa; Monika Johal; Ahran D Arnold; Nadine Ali; Khulat A Saqi; Katherine March; Leah Burden; Darrel P Francis; Zachary I Whinnett; Matthew J Shun-Shin
Journal:  Pacing Clin Electrophysiol       Date:  2022-01-28       Impact factor: 1.912

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.