BD rapid serum tubes reduce false positive plasma troponin T results on the Roche Cobas e411 analyzer.

Christopher D Koch1, Amy M Wockenfus, Amy K Saenger, Allan S Jaffe, Brad S Karon.   

Abstract

OBJECTIVES: In an attempt to reduce false positive results and improve turnaround time, we investigated the BD Rapid Serum Tube as an alternate sample type to lithium heparin plasma for Roche Troponin T analysis on the Roche Cobas e411 analyzer. DESIGN AND METHODS: BD Plasma Separator Tubes (PST) and Rapid Serum Tubes (RST) were collected in tandem from Emergency Department patients who had clinical orders for Troponin T over a 1 month period.
RESULTS: RST and PST samples yielded analytically and clinically concordant Troponin T results on the Roche Cobas e411. Rare false positive results in lithium heparin samples were not observed with rapid clot serum tubes.
CONCLUSIONS: RST samples are appropriate for stat Troponin T analysis, and appear to reduce the incidence of rare false positive Troponin T results obtained with lithium heparin samples.
Copyright © 2012 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22561083     DOI: 10.1016/j.clinbiochem.2012.04.014

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281