Characterization of a novel two dimensional diode array the "magic plate" as a radiation detector for radiation therapy treatment.

J H D Wong1, I Fuduli, M Carolan, M Petasecca, M L F Lerch, V L Perevertaylo, P Metcalfe, A B Rosenfeld.   

Abstract

PURPOSE: Intensity modulated radiation therapy (IMRT) utilizes the technology of multileaf collimators to deliver highly modulated and complex radiation treatment. Dosimetric verification of the IMRT treatment requires the verification of the delivered dose distribution. Two dimensional ion chamber or diode arrays are gaining popularity as a dosimeter of choice due to their real time feedback compared to film dosimetry. This paper describes the characterization of a novel 2D diode array, which has been named the "magic plate" (MP). It was designed to function as a 2D transmission detector as well as a planar detector for dose distribution measurements in a solid water phantom for the dosimetric verification of IMRT treatment delivery.
METHODS: The prototype MP is an 11 × 11 detector array based on thin (50 μm) epitaxial diode technology mounted on a 0.6 mm thick Kapton substrate using a proprietary "drop-in" technology developed by the Centre for Medical Radiation Physics, University of Wollongong. A full characterization of the detector was performed, including radiation damage study, dose per pulse effect, percent depth dose comparison with CC13 ion chamber and build up characteristics with a parallel plane ion chamber measurements, dose linearity, energy response and angular response.
RESULTS: Postirradiated magic plate diodes showed a reproducibility of 2.1%. The MP dose per pulse response decreased at higher dose rates while at lower dose rates the MP appears to be dose rate independent. The depth dose measurement of the MP agrees with ion chamber depth dose measurements to within 0.7% while dose linearity was excellent. MP showed angular response dependency due to the anisotropy of the silicon diode with the maximum variation in angular response of 10.8% at gantry angle 180°. Angular dependence was within 3.5% for the gantry angles ± 75°. The field size dependence of the MP at isocenter agrees with ion chamber measurement to within 1.1%. In the beam perturbation study, the surface dose increased by 12.1% for a 30 × 30 cm(2) field size at the source to detector distance (SDD) of 80 cm whilst the transmission for the MP was 99%.
CONCLUSIONS: The radiation response of the magic plate was successfully characterized. The array of epitaxial silicon based detectors with "drop-in" packaging showed properties suitable to be used as a simplified multipurpose and nonperturbing 2D radiation detector for radiation therapy dosimetric verification.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22559625     DOI: 10.1118/1.3700234

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Process-based quality management for clinical implementation of adaptive radiotherapy.

Authors:  Camille E Noel; Lakshmi Santanam; Parag J Parikh; Sasa Mutic
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

2.  Real-time beam monitoring for error detection in IMRT plans and impact on dose-volume histograms : A multi-center study.

Authors:  Livia Marrazzo; Chiara Arilli; Marlies Pasler; Martijn Kusters; Richard Canters; Luca Fedeli; Silvia Calusi; Marta Casati; Cinzia Talamonti; Gabriele Simontacchi; Lorenzo Livi; Stefania Pallotta
Journal:  Strahlenther Onkol       Date:  2017-12-18       Impact factor: 3.621

3.  X-Ray Beam Segment Size and Entrance Location Effects on the Integral Quality Monitor (IQM®) Signal and Usefulness in Predicting Complex Segment Output Signals.

Authors:  Mahuvava C; Du Plessis F C P
Journal:  J Biomed Phys Eng       Date:  2020-08-01

4.  Sensitivity of the IQM and MatriXX detectors in megavolt photon beams.

Authors:  Oluwaseyi Michael Oderinde; Freek du Plessis
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-02

5.  HDR brachytherapy in vivo source position verification using a 2D diode array: A Monte Carlo study.

Authors:  Joel Poder; Dean Cutajar; Susanna Guatelli; Marco Petasecca; Andrew Howie; Joseph Bucci; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-06-01       Impact factor: 2.102

6.  Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy.

Authors:  David Hoffman; Eunah Chung; Clayton Hess; Robin Stern; Stanley Benedict
Journal:  J Appl Clin Med Phys       Date:  2016-11-29       Impact factor: 2.102

7.  Characterization of MOSkin detector for in vivo skin dose measurement during megavoltage radiotherapy.

Authors:  Wei Loong Jong; Jeannie Hsiu Ding Wong; Ngie Min Ung; Kwan Hoong Ng; Gwo Fuang Ho; Dean L Cutajar; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

8.  Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate.

Authors:  Ziyad A Alrowaili; Michael L F Lerch; Marco Petasecca; Martin G Carolan; Peter E Metcalfe; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

9.  Diode-based transmission detector for IMRT delivery monitoring: a validation study.

Authors:  Taoran Li; Q Jackie Wu; Thomas Matzen; Fang-Fang Yin; Jennifer C O'Daniel
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  CyberKnife® fixed cone and Iris™ defined small radiation fields: Assessment with a high-resolution solid-state detector array.

Authors:  Giordano Biasi; Marco Petasecca; Susanna Guatelli; Ebert A Martin; Garry Grogan; Benjamin Hug; Jonathan Lane; Vladimir Perevertaylo; Tomas Kron; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-07-12       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.