Literature DB >> 22559612

Pilot program on patient dosimetry in pediatric interventional cardiology in Chile.

Carlos Ubeda1, Eliseo Vano, Patricia Miranda, Fernando Leyton.   

Abstract

PURPOSE: The aim of this study was to present the results of a pilot program on patient dosimetry carried out in Chile during the last 5 yr, using a biplane x-ray angiography system settled for pediatrics. This research was conducted in Latin America under the auspices of the International Atomic Energy Agency (IAEA) supporting programs on radiological protection (RP) of patients.
METHODS: Patient age, gender, weight, height, number of cine series, total number of cine frames, fluoroscopy time, and two dosimetric quantities [air kerma-area product (P(ka)) and cumulative dose (CD) at the patient entrance reference point] were recorded for each procedure.
RESULTS: The study includes 544 patients grouped into four age groups. The distributions by age group were 150 for <1 yr; 203 for 1 to <5 yr; 97 for 5 to <10 yr; and 94 for 10 to <16 yr. Median values of P(ka) and CD for the four age groups were 0.94, 1.46, 2.13, and 5.03 Gy cm(2) and 23.9, 26.8, 33.5, and 51.6 mGy, respectively. No significant statistical differences were found between diagnostic and therapeutic procedures. A moderate correlation (r = 0.64) was seen between P(ka) and patient weight.
CONCLUSIONS: The dose values reported in this paper were lower than those published in the previous work for the same age groups as a result of the optimization actions carried out by cardiologists and medical physicists with the support of the IAEA. Methodology and results will be used as a starting point for a wider survey in Chile and Latin America with the goal to obtain regional diagnostic reference levels as recently recommended by the International Commission on Radiological Protection for interventional procedures.

Entities:  

Mesh:

Year:  2012        PMID: 22559612     DOI: 10.1118/1.3702590

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Radiation Exposure by Three-Dimensional Rotational Angiography (3DRA) During Trans-catheter Melody Pulmonary Valve Procedures (TMPV) in a Pediatric Cardiac Catheterization Laboratory.

Authors:  Hoang H Nguyen; David T Balzer; Joshua J Murphy; Ramzi Nicolas; Shabana Shahanavaz
Journal:  Pediatr Cardiol       Date:  2016-07-25       Impact factor: 1.655

2.  Standardizing Radiation Exposure during Cardiac Catheterization in Children with Congenital Heart Disease: Data from a Multicenter Brazilian Registry.

Authors:  João Luiz Manica; Vanessa Oliveira Duarte; Marcelo Ribeiro; Adam Hartley; Ricardo Petraco; Carlos Pedra; Raul Rossi
Journal:  Arq Bras Cardiol       Date:  2020-12       Impact factor: 2.000

3.  Eye lens radiation exposure in paediatric interventional treatment of retinoblastoma.

Authors:  A Obesso; L Alejo; C Huerga; F Sánchez-Muñoz; E Corredoira; A Fernández-Prieto; R Frutos; B Marín; G Garzón; J Peralta; C Ubeda; E Guibelalde
Journal:  Sci Rep       Date:  2019-12-27       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.