Literature DB >> 22558497

Comparison of loss of resistance technique between Epidrum® and conventional method for identifying the epidural space.

Seon Wook Kim1, Young Mi Kim, Soo Hwan Kim, Mi Hwa Chung, Young Ryong Choi, Eun Mi Choi.   

Abstract

BACKGROUND: Epidrum® is a recently developed, air operated, loss of resistance (LOR) device for identifying the epidural space. We investigated the usefulness of Epidrum® by comparing it with the conventional LOR technique for identifying the epidural space.
METHODS: One hundred eight American Society of Anesthesiologists (ASA) physical status I or II patients between the ages of 17 and 68 years old and who were scheduled for elective surgery under combined spinal-epidural anesthesia were enrolled in this study. The patients were randomized into two groups: one group received epidural anesthesia by the conventional LOR technique (C group) and the second group received epidural anesthesia using Epidrum® (ED group). While performing epidural anesthesia, the values of variables were recorded, including the number of failures, more than 2 attempts, the incidence of dural puncture, the time needed to locate the epidural space, the distance from the skin to the epidural space and ease of performance, and the satisfaction scores.
RESULTS: The ED group showed a lower failure rate, fewer cases of more than 2 attempts, a lesser time to identify the epidural space, and better ease and satisfaction scores of procedure than the C group, with statistical significance.
CONCLUSIONS: Using Epidrum® compared to the conventional LOR technique is an easy, rapid, and reliable method for identifying the epidural space.

Entities:  

Keywords:  Epidrum®; Epidural anesthesia; Loss of resistance

Year:  2012        PMID: 22558497      PMCID: PMC3337377          DOI: 10.4097/kjae.2012.62.4.322

Source DB:  PubMed          Journal:  Korean J Anesthesiol        ISSN: 2005-6419


  16 in total

1.  The role of fluoroscopy in cervical epidural steroid injections: an analysis of contrast dispersal patterns.

Authors:  Milan P Stojanovic; To-Nhu Vu; Onassis Caneris; Jan Slezak; Steven P Cohen; Christine N Sang
Journal:  Spine (Phila Pa 1976)       Date:  2002-03-01       Impact factor: 3.468

Review 2.  Mechanisms and management of an incomplete epidural block for cesarean section.

Authors:  Dmitry Portnoy; Rakesh B Vadhera
Journal:  Anesthesiol Clin North Am       Date:  2003-03

3.  A pilot study to compare the Episure Autodetect syringe with the glass syringe for identification of the epidural space in parturients.

Authors:  Ashraf S Habib; Ronald B George; Terrence K Allen; Adeyemi J Olufolabi
Journal:  Anesth Analg       Date:  2008-02       Impact factor: 5.108

4.  Anaesthesia for emergency caesarean section.

Authors:  B M Morgan; V Magni; T Goroszenuik
Journal:  Br J Obstet Gynaecol       Date:  1990-05

5.  A change in resistance? A survey of epidural practice amongst obstetric anaesthetists.

Authors:  T K Howell; D P Prosser; M Harmer
Journal:  Anaesthesia       Date:  1998-03       Impact factor: 6.955

6.  Incidence of epidural catheter replacement in parturients: a retrospective chart review.

Authors:  S Eappen; A Blinn; S Segal
Journal:  Int J Obstet Anesth       Date:  1998-10       Impact factor: 2.603

7.  Success of spinal and epidural labor analgesia: comparison of loss of resistance technique using air versus saline in combined spinal-epidural labor analgesia technique.

Authors:  Lydia S Grondin; Kenneth Nelson; Vernon Ross; Orlando Aponte; Sherman Lee; Peter H Pan
Journal:  Anesthesiology       Date:  2009-07       Impact factor: 7.892

8.  Continuous epidural infusion of morphine versus single epidural injection of extended-release morphine for postoperative pain control after arthroplasty: a retrospective analysis.

Authors:  Stephanie Vanterpool; Randall Coombs; Karamarie Fecho
Journal:  Ther Clin Risk Manag       Date:  2010-06-24       Impact factor: 2.423

9.  The Episure syringe: a novel loss of resistance syringe for locating the epidural space.

Authors:  Edward T Riley; Brendan Carvalho
Journal:  Anesth Analg       Date:  2007-10       Impact factor: 5.108

10.  Loss of resistance technique for paravertebral nerve blockade using the Episure Autodetect Syringe--a case report.

Authors:  Derick A Mundey; Chester C Buckenmaier; Anthony R Plunkett
Journal:  Pain Med       Date:  2009-05-04       Impact factor: 3.750

View more
  3 in total

1.  Raman Spectroscopy Differentiates Each Tissue from the Skin to the Spinal Cord: A Novel Method for Epidural Needle Placement?

Authors:  T Anthony Anderson; Jeon Woong Kang; Tatyana Gubin; Ramachandra R Dasari; Peter T C So
Journal:  Anesthesiology       Date:  2016-10       Impact factor: 7.892

Review 2.  Localization of epidural space: A review of available technologies.

Authors:  Hesham Elsharkawy; Abraham Sonny; Ki Jinn Chin
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2017 Jan-Mar

3.  Feasibility of Using the Epidural Space Detecting Device (EPI-DetectionTM) for Interlaminar Cervical Epidural Injection.

Authors:  Jiin Kang; Sam Sun Park; Chul Hwan Kim; Eui Chul Kim; Hyung Cheol Kim; Hyungseok Jeon; Kyung Hyun Kim; Dong Ah Shin
Journal:  J Clin Med       Date:  2020-07-23       Impact factor: 4.241

  3 in total