"New-generation" pulse oximeters in extremely low-birth-weight infants: how do they perform in clinical practice?

Anne C van der Eijk1, Sandra Horsch, Paul H C Eilers, Jenny Dankelman, Bert J Smit.   

Abstract

The aim of this study was to evaluate the performance of "new-generation" pulse oximeters in extremely low-birth-weight ([ELBW] ≤ 1000 g) infants. In a prospective crossover observational study, the performance of pulse oximeters of 3 brands (Masimo, Nellcor, and Philips) was evaluated by dual SpO2 measurement in ELBW infants. Disposable probes of either equal or different brands were placed around both feet of the patient simultaneously for approximately 4 hours. Probes were switched between feet every hour. Absolute differences in SpO2 values (ΔSpO2) and the bias between brands were studied. Nine ELWB infants were included (gestational age: mean ± SD = 26(3)/7 ± 1 4/7 weeks). The median (range) ΔSpO2 was 2% (0%-26%). In 9% of the time, ΔSpO2 was 5% or more. The variance of the difference of the 3 pulse oximeter brands was not significantly different. No consequent bias between brands was found. Simultaneously obtained pulse oximeter measurements from the feet of ELBW infants differ from each other. Our results suggest that it is not the brand but the handling of the pulse oximeter in clinical practice, such as the place and positioning of the probe, that influences the performance of the pulse oximeter the most. Improvement in the accuracy of oxygen-monitoring techniques for ELBW infants is required.

Entities:  

Mesh:

Year:  2012        PMID: 22551866     DOI: 10.1097/JPN.0b013e31825277bd

Source DB:  PubMed          Journal:  J Perinat Neonatal Nurs        ISSN: 0893-2190            Impact factor:   1.638


  2 in total

1.  Performance of Two Different Pulse Oximeters in Neonatal Transition.

Authors:  Sunita Jeyapal; Peter Prasanth Kumar Kommu; M Manikandan; Lalitha Krishnan
Journal:  Indian J Pediatr       Date:  2016-08-19       Impact factor: 1.967

2.  MEchatronic REspiratory System SImulator for Neonatal Applications (MERESSINA) project: a novel bioengineering goal.

Authors:  Rosa T Scaramuzzo; Massimiliano Ciantelli; Ilaria Baldoli; Lisa Bellanti; Marzia Gentile; Francesca Cecchi; Emilio Sigali; Selene Tognarelli; Paolo Ghirri; Stefano Mazzoleni; Arianna Menciassi; Armando Cuttano; Antonio Boldrini; Cecilia Laschi; Paolo Dario
Journal:  Med Devices (Auckl)       Date:  2013-08-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.