L P J Teunissen1, A de Haan, J J de Koning, H A M Daanen.
Abstract
Core temperature measurement with an ingestible telemetry pill has been scarcely investigated during extreme rates of temperature change, induced by short high-intensity exercise in the heat. Therefore, nine participants performed a protocol of rest, (sub)maximal cycling and recovery at 30 °C. The pill temperature (T(pill)) was compared with the rectal temperature (T(re)) and esophageal temperature (T(es)). T(pill) corresponded well to T(re) during the entire trial, but deviated considerably from T(es) during the exercise and recovery periods. During maximal exercise, the average ΔT(pill)-T(re) and ΔT(pill)-T(es) were 0.13 ± 0.26 and -0.57 ± 0.53 °C, respectively. The response time from the start of exercise, the rate of change during exercise and the peak temperature were similar for T(pill) and T(re.) T(es) responded 5 min earlier, increased more than twice as fast and its peak value was 0.42 ± 0.46 °C higher than T(pill). In conclusion, also during considerable temperature changes at a very high rate, T(pill) is still a representative of T(re). The extent of the deviation in the pattern and peak values between T(pill) and T(es) (up to >1 °C) strengthens the assumption that T(pill) is unsuited to evaluate central blood temperature when body temperatures change rapidly.Entities: