Sanghoon Park1, Hoon Jai Chun, Bora Keum, Yeon Seok Seo, Yong Sik Kim, Yoon Tae Jeen, Hong Sik Lee, Soon Ho Um, Chang Duck Kim, Ho Sang Ryu.
Abstract
Purpose. PillCam SB2 capsule endoscopy, an upgraded version of widely used SB capsule endoscopy, was examined for its performance by comparing with SB. Methods. Examinees with various indications were enrolled for SB2 capsule endoscopy; subjects were also enlisted for the old SB capsule endoscopy. Number of photo images containing papilla of Vater was counted. Shape of the papilla seen in each image was evaluated by scoring 3 (fully observable papilla), 2 (more than half outline), or 1 (less than half outline) points. Images obtained from SB and SB2 were also subjectively compared; resolution and brightness were scored by six experienced endoscopists. Results. Baseline characteristics of two study groups (n = 30 each) were not significantly different. Number of images of the papilla revealed to show similar results between SB (3.1 ± 1.1, range 1~5) and SB2 (3.1 ± 1.5, range 1~8) (P = 0.62). The maximum points of outline of papilla evaluated from each subject were also similar between two groups. New SB2 revealed to be superior to SB in terms of resolution but not significantly different in brightness. Conclusion. Our study showed that superiority of SB2 over SB is rather marginal on examining duodenal papilla.Entities:
Year: 2012 PMID: 22548051 PMCID: PMC3324898 DOI: 10.1155/2012/202935
Source DB: PubMed Journal: Gastroenterol Res Pract ISSN: 1687-6121 Impact factor: 2.260
Baseline characteristics and indications for wireless capsule endoscopy.
| SB ( | SB2 ( |
| |
|---|---|---|---|
| Sex (M/F) | 14/16 | 15/15 | NS |
| Age | 46.1 ± 10.1 | 51.4 ± 9.2 | NS |
| Healthy volunteers | 3 | 2 | NS |
| Obscure GIB | 18 | 19 | NS |
| Abdominal pain | 5 | 6 | NS |
| Chronic diarrhea | 4 | 3 | NS |
(SB: small bowel capsule endoscope, SB2: small bowel capsule endoscope version 2, GIB: gastrointestinal bleeding, and NS: not significant)
*Values shown as mean ± SD or No.
Results of wireless capsule endoscopy comparing SB and SB2 capsule.
| SB | SB2 |
| |
|---|---|---|---|
| Detection rate (%) | 43.3 (13/30) | 50.0 (15/30) | 0.796 |
| No. of images of papilla | 3.1 ± 1.1 | 3.1 ± 1.5 | 0.571 |
| Image grading (max) | |||
| 3 points | 4/13 | 4/15 | |
| 2 points | 5/13 | 7/15 | |
| 1 point | 4/13 | 4/15 |
*Values shown as mean ± SD or No. max: maximum.
Figure 1Representative images and scorings of papillary outline (marked on bottom right) of each capsule endoscopy (upper and lower panel for SB2 and SB, resp.). The octangular outline of SB2 images were hidden with black-colored paper in order not to be discerned by experienced gastroenterologists.
Subjective scoring of resolution and brightness on SB and SB2 groups.
| Parameters | SB ( | SB2 ( |
|
|---|---|---|---|
| Resolution | 0.93 ± 0.62 | 1.55 ± 0.40 | 0.001 |
| Brightness | 1.21 ± 0.70 | 1.57 ± 0.51 | 0.165 |
*Values shown as mean ± SD or No.