Literature DB >> 22545887

Valiant thoracic stent-graft deployed with the new captivia delivery system: procedural and 30-day results of the Valiant Captivia registry.

Robin H Heijmen1, Matt M Thompson, Rossella Fattori, Yigit Goktay, Omke E Teebken, Karl Heinz Orend.   

Abstract

PURPOSE: To evaluate procedural and 30-day outcomes of thoracic endovascular aortic repair (TEVAR) employing the Valiant Thoracic Stent Graft with the Captivia Delivery System.
METHODS: Enrollment in the study ( www.ClinicalTrials.com identifier NCT01181947) included all eligible patients implanted with the Valiant Captivia System retrospectively and prospectively at 15 sites in Europe and Turkey between October 2009 and June 2010. In the 100 treated patients (81 men; mean age 64.6 ± 12.0 years, range 25-87), indications included descending thoracic aortic aneurysm (TAA, 49.0%) and aortic dissection (42.0%).
RESULTS: Technical success was 100.0%, with no misaligned deployments or aortic perforations. Mean follow-up was 68.9 ± 34.9 days (range 20-147, median 61). The 30-day rate of all-cause mortality was 4.0% (all 4 cases procedure-related, 3 device-related). Retrograde type A dissection occurred in 2 patients. The only conversion to open surgery was successful in a patient experiencing intraoperative aneurysm rupture. Stroke occurred in 4 (4.0%) patients and paraplegia in 1 (1.0%). Among 66 patients with 30-day imaging studies evaluable for endoleak, 4 (6.1%) had type I and 7 (10.6%) had type II endoleak; there were no types III or IV. Within 30 days, no secondary endovascular procedures were required due to endoleak. One patient with type II endoleak died 3 weeks postimplantation before scheduled embolization.
CONCLUSION: In this analysis of procedural and 30-day results, the high technical success and clinical outcome rates showed that the Valiant Thoracic Stent Graft with the new Captivia Delivery System has promising capacity to treat a variety of thoracic aortic conditions in a range of anatomies.

Entities:  

Mesh:

Year:  2012        PMID: 22545887     DOI: 10.1583/11-3652MR.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  6 in total

Review 1.  Emergent aortic surgery in octogenarians: is the advanced age a contraindication?

Authors:  Mario Castaño; Javier Gualis; Jose M Martínez-Comendador; Elio Martín; Pasquale Maiorano; Laura Castillo
Journal:  J Thorac Dis       Date:  2017-05       Impact factor: 2.895

2.  Impact of Female Sex on Outcomes of Patients Undergoing Thoracic Endovascular Aortic Aneurysm Repair: A Ten-Year Retrospective Nationwide Study in France.

Authors:  Fabien Lareyre; Juliette Raffort; Christian-Alexander Behrendt; Arindam Chaudhuri; Cong Duy Lê; Roxane Fabre; Christian Pradier; Laurent Bailly
Journal:  J Clin Med       Date:  2022-04-18       Impact factor: 4.964

3.  Red Blood Cell Distribution Width: A Prognostic Marker in Patients With Type B Aortic Dissection Undergoing Endovascular Aortic Repair.

Authors:  Cheng Jiang; Anbang Liu; Lei Huang; Quanjun Liu; Yuan Liu; Qingshan Geng
Journal:  Front Cardiovasc Med       Date:  2022-05-02

4.  Endovascular treatment of complex diseases of the thoracic aorta-10 years single centre experience.

Authors:  Piotr Buczkowski; Mateusz Puślecki; Natalia Majewska; Tomasz Urbanowicz; Marcin Misterski; Robert Juszkat; Jerzy Kulesza; Bartosz Żabicki; Sebastian Stefaniak; Marcin Ligowski; Lukasz Szarpak; Marek Jemielity; Eva Rivas; Kurt Ruetzler; Bartłomiej Perek
Journal:  J Thorac Dis       Date:  2019-06       Impact factor: 2.895

5.  A Right Intercosto-Bronchial Artery Derived from the Thyrocervical Trunk: An Unusual Cause of Type II Endoleak Post Thoracic Aortic Stenting.

Authors:  Andrew Mtl Choong; Ken Mitchell; Jason Jenkins
Journal:  Korean J Thorac Cardiovasc Surg       Date:  2015-04-05

Review 6.  Endovascular Repair in Acute Complicated Type B Aortic Dissection: 3-Year Results from the Valiant US Investigational Device Exemption Study.

Authors:  Chang Young Lim
Journal:  Korean J Thorac Cardiovasc Surg       Date:  2017-06-05
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.