Literature DB >> 22538250

Are conventional reconstruction plates equivalent to precontoured locking plates for distal humerus fracture fixation? A biomechanics cadaver study.

Ryan C Koonce1, Todd H Baldini, Steven J Morgan.   

Abstract

BACKGROUND: The optimal plate type and configuration for distal humerus fracture fixation has yet to be defined. Available biomechanical studies show conflicting results. No existing studies compare conventional reconstruction plates to newer precontoured distal humerus locking plates in both parallel and perpendicular configurations.
METHODS: Three groups of humerus specimens were compared via biomechanical testing in a cadaver model simulating metaphyseal comminution. Group 1 consisted of conventional reconstruction plates in a perpendicular configuration. Group 2 used precontoured locking plates in a perpendicular configuration. Group 3 used precontoured locking plates in a parallel configuration. Each group was tested for stiffness in anterior bending, posterior bending, axial compression, and torsion. The specimens then underwent cyclic loading followed by single load to failure in posterior bending.
FINDINGS: There was no significant difference between the three groups for anterior bending, posterior bending, axial compression, or torsional stiffness. There was no significant difference in load to failure for any of the three groups. Screw loosening was significantly higher in Group 1 when compared to Groups 2 and 3 after cyclic loading.
INTERPRETATION: In the early postoperative period, less expensive perpendicular conventional reconstruction plate constructs provide similar stiffness and load to failure properties to newer precontoured locking plate systems regardless of plate configuration.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22538250     DOI: 10.1016/j.clinbiomech.2012.03.008

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  4 in total

1.  Parallel versus orthogonal plate osteosynthesis of adult distal humerus fractures: a meta-analysis of biomechanical studies.

Authors:  Chien-An Shih; Wei-Ren Su; Wei-Chin Lin; Tai-Wei Tai
Journal:  Int Orthop       Date:  2018-04-20       Impact factor: 3.075

2.  Evaluation of risk factors for stiffness after distal humerus plating.

Authors:  Onur Tunalı; Ali Erşen; Tuna Pehlivanoğlu; Serkan Bayram; Ata C Atalar; Mehmet Demirhan
Journal:  Int Orthop       Date:  2018-02-01       Impact factor: 3.075

3.  Modified Tension Band Wiring in Adult Distal Humeral Fracture Types A2 and C1.

Authors:  Saeed Kokly; Alessandro Castagna; Mohamadreza Mohamadreza
Journal:  Arch Bone Jt Surg       Date:  2019-07

4.  Biomechanical comparison of orthogonal versus parallel double plating systems in intraarticular distal humerus fractures.

Authors:  Ata C Atalar; Onur Tunalı; Ali Erşen; Mehmet Kapıcıoğlu; Yavuz Sağlam; Mehmet S Demirhan
Journal:  Acta Orthop Traumatol Turc       Date:  2016-12-10       Impact factor: 1.511

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.