Literature DB >> 22535478

Repeatability of edited lactate and other metabolites in astrocytoma at 3T.

Mary A McLean1, Amy Sun, Thomas E Bradstreet, Andrea K Schaeffer, Haiying Liu, Robert Iannone, Gary Herman, Radha A Railkar, Ilse Joubert, Jonathan H Gillard, Stephen J Price, John R Griffiths.   

Abstract

PURPOSE: To assess the repeatability of measurement of lactate and other metabolites in tumors using magnetic resonance spectroscopy (MRS).
MATERIALS AND METHODS: MRS with spectral editing for lactate was performed on 10 patients with astrocytoma (two Grade III, eight Grade IV) using an 8-channel receive coil at 3T. Lactate, lipid, choline, creatine, and N-acetyl aspartate (NAA) signals were measured in regions of tumor and contralateral white matter. Metabolites were quantified relative to unsuppressed water using LCModel fitting software.
RESULTS: The within-patient coefficients of variation were ≈16% (tumor lactate), 6%-8% (tumor choline and contralateral choline, creatine, and NAA), and 22% (tumor lipid). As expected due to their low concentration in normal tissue, lactate and lipid were not reliably detected in white matter but were found at high levels in most tumors. NAA and creatine were lower in tumors than in normal white matter, and choline varied between above- and below-normal values. No consistent short-term variation in metabolite levels was observed, despite differences in the time elapsed since administration of contrast agent.
CONCLUSION: MRS appears repeatable enough to provide longitudinal measures of metabolite content in tumors and contralateral tissue in the brain in vivo.
Copyright © 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22535478     DOI: 10.1002/jmri.23673

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

Review 1.  Edited 1 H magnetic resonance spectroscopy in vivo: Methods and metabolites.

Authors:  Ashley D Harris; Muhammad G Saleh; Richard A E Edden
Journal:  Magn Reson Med       Date:  2017-02-02       Impact factor: 4.668

2.  Utility of multiparametric 3-T MRI for glioma characterization.

Authors:  Bhaswati Roy; Rakesh K Gupta; Andrew A Maudsley; Rishi Awasthi; Sulaiman Sheriff; Meng Gu; Nuzhat Husain; Sudipta Mohakud; Sanjay Behari; Chandra M Pandey; Ram K S Rathore; Daniel M Spielman; Jeffry R Alger
Journal:  Neuroradiology       Date:  2013-02-02       Impact factor: 2.804

3.  Single-shot single-voxel lactate measurements using FOCI-LASER and a multiple-quantum filter.

Authors:  Geoffrey S Payne; Nandita M deSouza; Christina Messiou; Martin O Leach
Journal:  NMR Biomed       Date:  2015-04       Impact factor: 4.044

4.  Quantifying the CDK inhibitor VMY-1-103's activity and tissue levels in an in vivo tumor model by LC-MS/MS and by MRI.

Authors:  Paul Sirajuddin; Sudeep Das; Lymor Ringer; Olga C Rodriguez; Angiela Sivakumar; Yi-Chien Lee; Aykut Üren; Stanley T Fricke; Brian Rood; Alpay Ozcan; Sean S Wang; Sana Karam; Venkata Yenugonda; Patricia Salinas; Emanuel Petricoin; Michael Pishvaian; Michael P Lisanti; Yue Wang; Richard Schlegel; Bahram Moasser; Chris Albanese
Journal:  Cell Cycle       Date:  2012-09-14       Impact factor: 4.534

  4 in total