Literature DB >> 22534758

Genotyping from saliva with a one-step microdevice.

Ilija Pjescic1, Niel Crews.   

Abstract

This paper presents a disposable microfluidic device for on-chip lysing, PCR, and analysis in one continuous-flow process. Male-female sex determination was performed with human saliva in less than 20 min from spit to finish, and requiring only seconds of manual sample handling. This genetic analysis was based on the amplification and detection of the DYZ1 repeat region unique to the Y-chromosome. The flow-through microfluidic chip consisted of a single serpentine channel designed to guide samples through 42 heating and cooling cycles. Cycling was performed by matching the local channel geometry to a steady-state temperature gradient established across the microfluidic chip. 38 channel segments were designed for rapid low volume PCR, and four were optimized for spatial DNA melting analysis. Fluorescence detection was used to monitor the amplification and to capture the melting signature of the amplicon was performed with a basic 8-bit CCD camera. The microfluidic device itself was fabricated from microscope slides and a double-sided tape. The simplicity of the system and its robust performance combine in an elegant solution for lab-on-a-chip genetic analysis.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22534758     DOI: 10.1039/c2lc00010e

Source DB:  PubMed          Journal:  Lab Chip        ISSN: 1473-0189            Impact factor:   6.799


  2 in total

1.  A microfluidic chip integrating DNA extraction and real-time PCR for the detection of bacteria in saliva.

Authors:  Emily A Oblath; W Hampton Henley; Jean Pierre Alarie; J Michael Ramsey
Journal:  Lab Chip       Date:  2013-04-07       Impact factor: 6.799

2.  A newly developed paper embedded microchip based on LAMP for rapid multiple detections of foodborne pathogens.

Authors:  Mimi Zhang; Jinfeng Liu; Zhiqiang Shen; Yongxin Liu; Yang Song; Yu Liang; Zhende Li; Lingmei Nie; Yanjun Fang; Youquan Zhao
Journal:  BMC Microbiol       Date:  2021-06-28       Impact factor: 3.605

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.