Literature DB >> 22534691

Reduction of pullout strength caused by reinsertion of 3.5-mm cortical screws.

Amir Matityahu1, Christof Hurschler, Markus Badenhop, Christina Stukenborg-Colsman, Hazibullah Waizy, Brock Wentz, Meir Marmor, Christian Krettek.   

Abstract

INTRODUCTION: Osteosynthesis of the tibia, tibial plafond, and calcaneus is commonly performed with plates and 3.5-mm self-tapping cortical screws. Screw insertion and reinsertion within the same hole in the bone may occur during surgery. Therefore, the purpose of this study was to evaluate the pullout strength of 3.5-mm self-tapping screws with up to 5 re-insertions in the diaphysis of the tibia, metaphysis of the distal tibia, calcaneus, and a polyurethane synthetic bone model.
METHODS: Screws were inserted into a synthetic bone model and 5 pairs of human cadaveric diaphyseal tibiae, distal tibiae, and calcanei. The bone was predrilled, and then 3.5-mm cortical self-tapping 316 L stainless steel screws with a washer were inserted bicortically. Screws were inserted from 1 to 5 times at each location. The screws were grasped and subjected to 5-mm/min tensional force via the biaxial material testing systems machine. Statistical significance was determined using a paired 2-tailed t test.
RESULTS: There was a significant difference in the pullout strength of the tibial diaphysis (1710 ± 550 N), tibial metaphysis (471 ± 266 N), and calcaneus (238 ± 90 N; P < 0.01). The tibial diaphysis pullout strength was 1710 ± 550 N for one insertion differing significantly relative to the groups with 4 (average 1030 ± 543 N, P = 0.004) or 5 (average 364 ± 209 N, P < 0.001) insertions. The tibial metaphyseal pullout strength for the single insertion group was 471 ± 266 N and differed significantly relative to the 3 (P = 0.026), 4 (P = 0.044), and 5 (P = 0.042) insertion groups. The calcaneal pullout strength for the single insertion group was 238 ± 90 N with a significant difference of the 1, 3, and 4, versus the 5 insertion group (P = 0.027, 0.040, and 0.033, respectively). The synthetic bone model pullout strength decreased significantly from the one insertion group relative to all other insertion groups (group 1, 1167 ± 263 N; group 2, 768 ± 199 N; group 3, 694 ± 295 N; group 4, 662 ± 356 N; and group 5, 154 ± 183 N; P < 0.02).
CONCLUSIONS: There is a significant decrease in relative pullout strength of 3.5-mm self-tapping cortical screws when comparing the tibial diaphysis, tibial metaphysis, and calcaneus. There is also a significant decrease in 3.5-mm self-tapping screw pullout strength after repeated reinsertions in the synthetic bone model, mid-shaft tibia, metaphyseal tibia, and calcaneus. We recommend that during osteosynthesis, careful screw insertion, and minimal reinsertion be performed.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 22534691     DOI: 10.1097/BOT.0b013e31825490b1

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  5 in total

1.  Pullout Strength After Multiple Reinsertions in Radial Bone Fixation.

Authors:  Alfonso Mejia; Giovanni Solitro; Elena Gonzalez; Amit Parekh; Mark Gonzalez; Farid Amirouche
Journal:  Hand (N Y)       Date:  2018-09-06

2.  Kirschner wire prepared pilot holes improve screw pullout strength in synthetic osteoporotic-type bone.

Authors:  Hrayr G Basmajian; Joseph N Liu; Travis Scudday; Seth T Campbell; Nirav H Amin
Journal:  J Clin Orthop Trauma       Date:  2019-08-22

3.  Helical blade compression failure occurred during PFNA implantation: A rare case and ingenious solution.

Authors:  Chaoqun Wang; Qingxian Wang
Journal:  Medicine (Baltimore)       Date:  2019-07       Impact factor: 1.817

4.  Factors Associated With the Accuracy of Depth Gauge Measurements.

Authors:  Pengcheng Liu; Joanna Xi Xiao; Chen Zhao; Xiaodong Li; Guantong Sun; Fei Yang; Xiaoqing Wang
Journal:  Front Surg       Date:  2022-01-13

5.  Cut Cortical Screw Purchase in Diaphyseal Bone: A Biomedical Study.

Authors:  Alexander C Wendling; Joel White; Benjamin J Cooper; Chad M Corrigan; Bradley R Dart
Journal:  Kans J Med       Date:  2022-02-09
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.