Injection of gadolinium contrast through pediatric central venous catheters: a safety study.

John M Moriarty1, Geoffrey L Kung, Yanerys Ramos, Abbas N Moghaddam, Daniel B Ennis, J Paul Finn.   

Abstract

BACKGROUND: Catheter rupture during CT angiography has prompted policies prohibiting the use of electronic injectors with peripherally inserted central venous catheters (PICCs) not only for CT but also for MRI. Consequently, many institutions mandate hand injection for MR angiography, limiting precision of infusion rates and durations of delivery.
OBJECTIVE: To determine whether electronic injection of gadolinium-based contrast media through a range of small-caliber, single-lumen PICCs would be safe without risk of catheter rupture over the range of clinical protocols and determine whether programmed flow rates and volumes were realized when using PICCs for contrast delivery.
MATERIALS AND METHODS: Experiments were performed and recorded using the Medrad Spectris Solaris EP MR Injection System. PICC sizes, contrast media and flow rates were based on common institutional protocols.
RESULTS: No catheters were damaged during any experiments. Mean difference between programmed and delivered volume was 0.07 ± 0.10 mL for all experiments. Reduced flow rates and prolonged injection durations were observed when the injector's pressure-limiting algorithm was triggered, only in protocols outside the clinical range.
CONCLUSION: PICCs commonly used in children can withstand in vitro power injection of gadolinium-based contrast media at protocols significantly above clinical levels.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22526282     DOI: 10.1007/s00247-012-2397-z

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  17 in total

1.  Estimates of the cancer risks from pediatric CT radiation are not merely theoretical: comment on "point/counterpoint: in x-ray computed tomography, technique factors should be selected appropriate to patient size. against the proposition".

Authors:  D J Brenner; C D Elliston; E J Hall; W E Berdon
Journal:  Med Phys       Date:  2001-11       Impact factor: 4.071

2.  Transport of contrast agents in contrast-enhanced magnetic resonance angiography.

Authors:  Liang-Der Jou; Joseph Rapp; D Saloner
Journal:  Magn Reson Imaging       Date:  2004-05       Impact factor: 2.546

Review 3.  Complications of vascular catheters in the neonatal intensive care unit.

Authors:  Jayashree Ramasethu
Journal:  Clin Perinatol       Date:  2008-03       Impact factor: 3.430

4.  Central venous catheter integrity during mechanical power injection of iodinated contrast medium.

Authors:  Douglas B Macha; Rendon C Nelson; Laurens E Howle; John W Hollingsworth; Sebastian T Schindera
Journal:  Radiology       Date:  2009-09-29       Impact factor: 11.105

5.  Estimated risks of radiation-induced fatal cancer from pediatric CT.

Authors:  D Brenner; C Elliston; E Hall; W Berdon
Journal:  AJR Am J Roentgenol       Date:  2001-02       Impact factor: 3.959

Review 6.  Advances in the diagnosis and management of central venous access device infections in children.

Authors:  Asha Bowen; Jonathan Carapetis
Journal:  Adv Exp Med Biol       Date:  2011       Impact factor: 2.622

7.  Safe use of power injectors with central and peripheral venous access devices for pediatric CT.

Authors:  S C Kaste; C W Young
Journal:  Pediatr Radiol       Date:  1996

8.  Venous thrombosis associated with peripherally inserted central catheters: a retrospective analysis of the Cleveland Clinic experience.

Authors:  Roy F Chemaly; Joaquin Barbara de Parres; Susan J Rehm; Karim A Adal; Michelle V Lisgaris; Debbie S Katz-Scott; Susan Curtas; Steven M Gordon; Ezra Steiger; Jeffrey Olin; David L Longworth
Journal:  Clin Infect Dis       Date:  2002-04-03       Impact factor: 9.079

9.  Lower extremity stepping-table magnetic resonance angiography with multilevel contrast timing and segmented contrast infusion.

Authors:  Mark D Morasch; Jeremy Collins; F Scott Pereles; James C Carr; Mark K Eskandari; William H Pearce; J Paul Finn
Journal:  J Vasc Surg       Date:  2003-01       Impact factor: 4.268

10.  Maximal flow rates possible during power injection through currently available PICCs: an in vitro study.

Authors:  Ari I Salis; Anthony Eclavea; Matthew S Johnson; Nilesh H Patel; Debie G Wong; Gerald Tennery
Journal:  J Vasc Interv Radiol       Date:  2004-03       Impact factor: 3.464

View more
  1 in total

Review 1.  Systematic review of the safety and efficacy of contrast injection via venous catheters for contrast-enhanced computed tomography.

Authors:  S B Buijs; M W Barentsz; M L J Smits; J W C Gratama; P E Spronk
Journal:  Eur J Radiol Open       Date:  2017-09-29
  1 in total