Literature DB >> 22522531

Effects of microchannel geometry on preconcentration intensity in microfluidic chips with straight or convergent-divergent microchannels.

Chia-Lin Chen1, Ruey-Jen Yang.   

Abstract

Preconcentration microfluidic devices are fabricated incorporating straight or convergent-divergent microchannels and hydrogel or Nafion membranes. Sample preconcentration is achieved utilizing concentration-polarization effects. The effects of the microchannel geometry on the preconcentration intensity are systematically examined. It is shown that for the preconcentrator with the straight microchannel, the time required to achieve a satisfactory preconcentration intensity increases with an increasing channel depth. For the convergent-divergent microchannel, the preconcentration intensity increases with a reducing convergent channel width. Comparing the preconcentration performance of the two different microchannel configurations, it is found that for an equivalent width of the main microchannel, the concentration effect in the convergent-divergent microchannel is faster than that in the straight microchannel.
© 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

Entities:  

Mesh:

Year:  2012        PMID: 22522531     DOI: 10.1002/elps.201100493

Source DB:  PubMed          Journal:  Electrophoresis        ISSN: 0173-0835            Impact factor:   3.535


  5 in total

1.  Capillarity ion concentration polarization for spontaneous biomolecular preconcentration mechanism.

Authors:  Yoonjee Oh; Hyomin Lee; Seok Young Son; Sung Jae Kim; Pilnam Kim
Journal:  Biomicrofluidics       Date:  2016-01-07       Impact factor: 2.800

2.  Preconcentration of diluted mixed-species samples following separation and collection in a micro-nanofluidic device.

Authors:  Yi-Ying Chen; Ping-Hsien Chiu; Chen-Hsun Weng; Ruey-Jen Yang
Journal:  Biomicrofluidics       Date:  2016-02-18       Impact factor: 2.800

3.  Ion concentration polarization on paper-based microfluidic devices and its application to preconcentrate dilute sample solutions.

Authors:  Ruey-Jen Yang; Hao-Hsuan Pu; Hsiang-Li Wang
Journal:  Biomicrofluidics       Date:  2015-02-18       Impact factor: 2.800

4.  Preconcentration and Separation of Mixed-Species Samples Near a Nano-Junction in a Convergent Microchannel.

Authors:  Ping-Hsien Chiu; Chen-Hsun Weng; Ruey-Jen Yang
Journal:  Sensors (Basel)       Date:  2015-12-05       Impact factor: 3.576

5.  Quantification of Vortex Generation Due to Non-Equilibrium Electrokinetics at the Micro/Nanochannel Interface: Particle Tracking Velocimetry.

Authors:  Seung Jun Lee; Kilsung Kwon; Tae-Joon Jeon; Sun Min Kim; Daejoong Kim
Journal:  Micromachines (Basel)       Date:  2016-07-21       Impact factor: 2.891

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.