Mohammad Bagher Gholivand1, Mir Mahdi Abolghasemi.
Abstract
Highly porous polyaniline/hexagonally ordered silica sorbent was used for fabrication of the inside needle capillary adsorption trap (INCAT) device. Polyaniline/SBA-15 nanocomposite was synthesized via chemical polymerization technique. The fabricated INCAT device was evaluated to the extraction of some polycyclic aromatic hydrocarbons (PAHs) from aqueous sample solutions in combination with gas chromatography (GC)-mass spectrometry (MS). A one at-the-time optimization strategy was applied for optimizing the important extraction parameters such as extraction temperature, extraction time, ionic strength, sampling flow rate, desorption time, and desorption temperature. In optimum conditions (extraction temperature 55 °C, extraction time 20 min, ionic strength 20% (w/v), flow rate 4.5 mL min(-1), desorption temperature 270 °C, desorption time 3 min) the repeatability for one INCAT device (n = 4), expressed as relative standard deviation, was between 4.2 and 10.2% for the tested compounds. The quantitation limits for the studied compounds were between 1 and 5 pg mL(-1). The developed method was successfully applied to spring water sample which was spiked with PAHs with the relative recovery percentages of 87.3-109.1%. The developed method offers the advantage of being simple to use, with shorter analysis times, lower cost of equipment, and thermal stability.Entities: