B Hou1, J H Easter, J A Nees, Z He, A G R Thomas, K Krushelnick.
Abstract
The multiphoton intrapulse interference phase scan (MIIPS) technique is modified to optimize the compressor settings of a chirped pulse amplification (CPA) laser system. Here, we use the compressor itself to perform the phase scan inherent in MIIPS measurement . A frequency-resolved optical gating measurement shows that the pulse duration of the compressor optimized using the modified MIIPS technique is 33.8 fs with a 2.24 rad temporal phase variation above 2% intensity. The measured time-bandwidth product is 0.60, which is close to that of transform-limited Gaussian pulse (0.44).Year: 2012 PMID: 22513694 DOI: 10.1364/OL.37.001385