Thoracic and abdominal aortic diameter measurement by MRI using plain axial volumetric interpolated breath-hold examination in epidemiologic research: a validation study.

Birger Mensel1, Katrin Hegenscheid, Lydia Hesselbarth, Michael Wenzel, Norbert Hosten, Ralf Puls.   

Abstract

RATIONALE AND
OBJECTIVES: This study evaluates the validity and reliability of measuring the diameters of the thoracic and abdominal aorta from plain volumetric interpolated breath-hold examination (VIBE) images.
MATERIALS AND METHODS: The study included 50 male subjects from the population-based Study of Health in Pomerania. They underwent imaging of the thoracic and abdominal aorta at 1.5 Tesla using a contrast-enhanced magnetic resonance angiography (CE-MRA) and plain VIBE sequence. Diameters were measured at five predefined anatomic sites from reformatted orthogonal CE-MRA images and axial plain VIBE images. The measurements were validated using Pearson correlation and Bland-Altman analysis. The Bland-Altman method was also used to assess reliability.
RESULTS: Comparison of the diameters measured from CE-MRA and VIBE images revealed strong correlation for the ascending, descending, suprarenal, and infrarenal aorta with r = 0.95 (P < .0001), r = 0.88 (P < .0001), 0.92 (P < .0001), and 0.87 (P < .0001), respectively. Measurement for the aortic arch was moderately correlated with r = 0.78 (P < .0001). Mean bias did not exceed 0.1 cm (6%). The 95% limits of agreement (LOA) were less than 0.5 cm (15%). Intra- and interobserver agreement showed a mean bias of less than 2%; the 95% LOA were less than 11%.
CONCLUSIONS: Axial measurement of the diameters of the thoracic and abdominal aorta using a plain axial VIBE sequence is highly valid and reliable, making it suitable for use in epidemiologic research.
Copyright © 2012 AUR. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2012        PMID: 22503891     DOI: 10.1016/j.acra.2012.03.012

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  3 in total

1.  Mandibular canal visibility using a plain volumetric interpolated breath-hold examination sequence in MRI.

Authors:  Chutamas Deepho; Hiroshi Watanabe; Junichiro Sakamoto; Tohru Kurabayashi
Journal:  Dentomaxillofac Radiol       Date:  2017-10-20       Impact factor: 2.419

2.  Thoracic and abdominal aortic diameters in a general population: MRI-based reference values and association with age and cardiovascular risk factors.

Authors:  Birger Mensel; Lydia Heßelbarth; Michael Wenzel; Jens-Peter Kühn; Marcus Dörr; Henry Völzke; Wolfgang Lieb; Katrin Hegenscheid; Roberto Lorbeer
Journal:  Eur Radiol       Date:  2015-07-25       Impact factor: 5.315

3.  High levels of thyroid-stimulating hormone are associated with aortic wall thickness in the general population.

Authors:  Till Ittermann; Roberto Lorbeer; Marcus Dörr; Tobias Schneider; Alexander Quadrat; Lydia Heßelbarth; Michael Wenzel; Ina Lehmphul; Josef Köhrle; Birger Mensel; Henry Völzke
Journal:  Eur Radiol       Date:  2016-03-14       Impact factor: 5.315