Literature DB >> 22503303

Proud osteochondral autograft versus synthetic plugs--contact pressures with cyclical loading in a bovine knee model.

Sameh Elguizaoui1, David C Flanigan, Joshua D Harris, Erin Parsons, Alan S Litsky, Robert A Siston.   

Abstract

PURPOSE: To determine whether proud synthetic and proud osteochondral plugs conform to native surrounding cartilage after cyclical loading and if there are differences in height and contact pressure after loading.
METHODS: Sixteen bovine knees were used. Each received one osteochondral plug and one of two types of synthetic plugs (Smith & Nephew TruFit® BGS plug or Kensey Nash OsseoFit® plug). Plugs were placed in the center of each condyle's highest pressure area of articulation identified prior to defect creation (control). Static loads of 800N were applied and contact pressures measured with Tekscan sensors.
RESULTS: Both types of synthetic grafts and the osteochondral grafts all subsided with cyclical loading (p<0.001). The OsseoFit® plug displayed a greater reduction in height than the TruFit® plug compared to osteochondral grafts. The OsseoFit® plugs had significantly lower height than osteochondral grafts after both periods of cyclical loading (p<0.001), while height of the TruFit® plugs was not significantly different than the heights of osteochondral grafts after either first (p=0.387) or second (p=0.261) periods of cyclical loading. Contact pressures for the OsseoFit® plugs were significantly lower than contact pressures for osteochondral grafts after both periods of cyclical loading (p<0.001 for both). There was no difference between the pressures of TruFit® and osteochondral plugs after the first (p=0.381) or second (p=0.292) periods of cyclical loading.
CONCLUSIONS: Precision and accuracy are demanded to achieve flush osteochondral plug placement and OsseoFit synthetic plugs may subside more with less pressure than TruFit plugs if placed proud.
Copyright © 2012 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22503303     DOI: 10.1016/j.knee.2012.03.008

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  4 in total

1.  Knotless anchors offer better prevention of meniscal excursion than knotted anchors: An experimental study of the bovine knee.

Authors:  Mehmet Burtaç Eren; Murat Aşçı; Ergin Tönük; Orhan Balta; Recep Kurnaz
Journal:  Acta Orthop Traumatol Turc       Date:  2020-01       Impact factor: 1.511

Review 2.  Cartilage regeneration.

Authors:  Rocky S Tuan; Antonia F Chen; Brian A Klatt
Journal:  J Am Acad Orthop Surg       Date:  2013-05       Impact factor: 3.020

3.  Tri-layered composite plug for the repair of osteochondral defects: in vivo study in sheep.

Authors:  Altug Yucekul; Deniz Ozdil; Nuri Hunkar Kutlu; Esra Erdemli; Halil Murat Aydin; Mahmut Nedim Doral
Journal:  J Tissue Eng       Date:  2017-04-13       Impact factor: 7.813

4.  Synthetic PVA Osteochondral Implants for the Knee Joint: Mechanical Characteristics During Simulated Gait.

Authors:  Tony Chen; Caroline Brial; Moira McCarthy; Russell F Warren; Suzanne A Maher
Journal:  Am J Sports Med       Date:  2021-08-04       Impact factor: 7.010

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.