Polymer-on-metal or metal-on-polymer total disc arthroplasty: does it make a difference?

Parshia M Moghadas1, Duncan E T Shepherd, David W L Hukins, Aziza Mahomed.   

Abstract

STUDY
DESIGN: Mechanical testing of total disc arthroplasty (TDA).
OBJECTIVE: To compare the friction between a polymer socket-on-metal ball and metal socket-on-polymer ball TDA. SUMMARY OF BACKGROUND DATA: A degenerate intervertebral disc can be replaced by TDA. The most common designs have a ball and socket articulation; the contact between the surfaces leads to friction. Friction needs to be minimized to prevent loosening and wear. One of the common material combinations in disc arthroplasty devices is the articulation of a metal socket on polymer ball. However, the combination of a polymer socket on metal ball (which is used in hip arthroplasty) has not been investigated for TDA.
METHODS: TDA models with either a polymer socket/metal ball or a metal socket/polymer ball were manufactured with ball radii of 10 and 14 mm, each with a radial clearance of 0.35 mm. Samples were tested using a spine simulator with a lubricant of diluted newborn calf serum. Each sample was subjected to an axial load of 1200 N; motions of flexion-extension, lateral bending, and axial rotation were then applied at frequencies of 0.25 to 2 Hz. Frictional torque was measured to compare the performance of the TDAs.
RESULTS: The frictional torque was found to be significantly higher for a disc with a metal socket/polymer ball than for a disc with a polymer socket/metal ball for both 10 and 14 mm radii in axial rotation, lateral bend, and extension. The frictional torque in flexion (0°-6°) was not found to be significantly different between the 2 different material combinations. However, when the flexion motion was reduced to 0° to 2°, frictional torque in the metal socket/polymer ball was found to be significantly higher than the polymer socket/metal ball.
CONCLUSION: TDA with a combination of a polymer socket/metal ball has lower friction than the conventional TDA with metal socket/polymer ball. This conclusion has implications in the design of TDA.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22498992     DOI: 10.1097/BRS.0b013e318257fdd3

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  4 in total

1.  Wear of the Charité® lumbar intervertebral disc replacement investigated using an electro-mechanical spine simulator.

Authors:  Parshia Moghadas; Aziza Mahomed; Duncan E T Shepherd; David W L Hukins
Journal:  Proc Inst Mech Eng H       Date:  2015-03       Impact factor: 1.617

2.  ISASS Policy Statement - Lumbar Artificial Disc.

Authors:  Jack Zigler; Rolando Garcia
Journal:  Int J Spine Surg       Date:  2015-03-12

Review 3.  Material Science in Cervical Total Disc Replacement.

Authors:  Martin H Pham; Vivek A Mehta; Alexander Tuchman; Patrick C Hsieh
Journal:  Biomed Res Int       Date:  2015-10-07       Impact factor: 3.411

4.  Differences in Tribological Behaviors upon Switching Fixed and Moving Materials of Tribo-pairs including Metal and Polymer.

Authors:  Aijie Xu; Pengyi Tian; Shizhu Wen; Fei Guo; Yueqiang Hu; Wenpeng Jia; Conglin Dong; Yu Tian
Journal:  Sci Rep       Date:  2017-10-12       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.