Alexander H Slocum1, Alexander H Slocum1, Joan E Spiegel.
Abstract
Endotracheal intubation is a frequently performed procedure in the prehospital setting, intensive care unit, and for patients undergoing surgery. The endotracheal tube cuff must be inflated to a pressure that prevents air leaks without compromising tracheal mucosal blood flow. For simultaneous endotracheal tube cuff inflation and measurement, we designed and tested a novel pressure-sensing syringe in vitro. The prototype was developed using a standard 10-mL polycarbonate syringe body that houses a plunger and a silicone rubber bellows, the pressure-sensing element. Bellow feasibility was determined and modeled using finite element analysis. Repeatability testing at each pressure measurement for each bellows (pressure versus deflection) was within an average standard deviation of 0.3 cm to 1.61 cm (1%-5% error). Using an aneroid manometer for comparison, there was excellent linear correlation with a Spearman rank of 0.99 (P < 0.001), up to 30 cm H(2)O.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22492187 DOI: 10.1213/ANE.0b013e31824abc4d