Effect of motion artefact on pulse oximeters: evaluation of four instruments and finger probes.

J A Langton1, C D Hanning.   

Abstract

The ability of the Ohmeda 3700, Nellcor N200, Datex Satlite Plus and Simed S100 pulse oximeters to detect induced hypoxaemia in the presence of motion artefact was assessed, under conditions of controlled vibration using an industrial vibration facility. Vibration at 4 Hz and 8 Hz induced increases in detection time for hypoxaemia and spurious decreases in the displayed SaO2 in some of the oximeters tested. Finger-dependent differences in oximeter performance and pulse rate registration were noted especially in those oximeters without ECG linkage (Ohmeda 3700 and Simed S100). Subsequently, eight different pulse oximeter finger probes were assessed for those characteristics that may predispose to motion artefact. There were marked differences in the mass of the probes, the forces exerted on the test finger and in the force required to displace the probes from the subject's finger. Differences in both the microprocessor programmes and the physical characteristics of the finger probes may explain the observed differences in function. Similar studies should form part of the standard evaluation of new pulse oximeters.

Entities:  

Mesh:

Year:  1990        PMID: 2248828     DOI: 10.1093/bja/65.4.564

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  10 in total

1.  Comparison of a prototype esophageal oximetry probe with two conventional digital pulse oximetry monitors in aortocoronary bypass patients.

Authors:  R C Prielipp; P E Scuderi; M H Hines; J L Atlee; J F Butterworth
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

2.  Exercise thermoregulatory responses following a 28-day sleep-high train-low regimen.

Authors:  Stylianos N Kounalakis; Ola Eiken; Igor B Mekjavic
Journal:  Eur J Appl Physiol       Date:  2012-03-10       Impact factor: 3.078

3.  Reflectance pulse oximetry at the forehead of newborns: the influence of varying pressure on the probe.

Authors:  A C Dassel; R Graaff; A Meijer; W G Zijlstra; J G Aarnoudse
Journal:  J Clin Monit       Date:  1996-11

4.  Identification of movement artefact by the Nellcor N-200 and N-3000 pulse oximeters.

Authors:  J L Plummer; A H Ilsley; R R Fronsko; H Owen
Journal:  J Clin Monit       Date:  1997-03

5.  Reflectance pulse oximetry at the forehead improves by pressure on the probe.

Authors:  A C Dassel; R Graaff; M Sikkema; A Meijer; W G Zijlstra; J G Aarnoudse
Journal:  J Clin Monit       Date:  1995-07

6.  Respiratory muscle endurance training: effect on normoxic and hypoxic exercise performance.

Authors:  Michail E Keramidas; Tadej Debevec; Mojca Amon; Stylianos N Kounalakis; Bostjan Simunic; Igor B Mekjavic
Journal:  Eur J Appl Physiol       Date:  2010-03       Impact factor: 3.078

7.  Nocturnal hypoxaemia after myocardial infarction: association with nocturnal myocardial ischaemia and arrhythmias.

Authors:  S Galatius-Jensen; J Hansen; V Rasmussen; J Bildsøe; M Therboe; J Rosenberg
Journal:  Br Heart J       Date:  1994-07

8.  A conversion formula for comparing pulse oximeter desaturation rates obtained with different averaging times.

Authors:  Jan Vagedes; Anja Bialkowski; Cornelia Wiechers; Christian F Poets; Klaus Dietz
Journal:  PLoS One       Date:  2014-01-28       Impact factor: 3.240

9.  Lightweight physiologic sensor performance during pre-hospital care delivered by ambulance clinicians.

Authors:  Alasdair J Mort; David Fitzpatrick; Philip M J Wilson; Chris Mellish; Anne Schneider
Journal:  J Clin Monit Comput       Date:  2015-03-25       Impact factor: 2.502

10.  Noninvasive Monitoring during Interhospital Transport of Newborn Infants.

Authors:  Georg M Schmölzer; Megan O'Reilly; Po-Yin Cheung
Journal:  Crit Care Res Pract       Date:  2013-02-24
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.