Literature DB >> 22482670

Computational modeling of shear-based hemolysis caused by renal obstruction.

Polina A Segalova1, K T Venkateswara Rao, Christopher K Zarins, Charles A Taylor.   

Abstract

As endovascular treatment of abdominal aortic aneurysms (AAAs) gains popularity, it is becoming possible to treat certain challenging aneurysmal anatomies with endografts relying on suprarenal fixation. In such anatomies, the bare struts of the device may be placed across the renal artery ostia, causing partial obstruction to renal artery blood flow. Computational fluid dynamics (CFD) was used to simulate blood flow from the aorta to the renal arteries, utilizing patient-specific boundary conditions, in three patient models and calculate the degree of shear-based blood damage (hemolysis). We used contrast-enhanced computed tomography angiography (CTA) data from three AAA patients who were treated with a novel endograft to build patient-specific models. For each of the three patients, we constructed a baseline model and endoframe model. The baseline model was a direct representation of the patient's 30-day post-operative CTA data. This model was then altered to create the endoframe model, which included a ring of metallic struts across the renal artery ostia. CFD was used to simulate blood flow, utilizing patient-specific boundary conditions. Pressures, flows, shear stresses, and the normalized index of hemolysis (NIH) were quantified for all patients. The overall differences between the baseline and endoframe models for all three patients were minimal, as measured though pressure, volumetric flow, velocity, and shear stress. The average NIH across the three baseline and endoframe models was 0.002 and 0.004, respectively. Results of CFD modeling show that the overall disturbance to flow caused by the presence of the endoframe struts is minimal. The magnitude of the NIH in all models was well below the accepted design and safety threshold for implantable medical devices that interact with blood flow.

Entities:  

Mesh:

Year:  2012        PMID: 22482670     DOI: 10.1115/1.4005850

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  3 in total

1.  Patient-specific changes in aortic hemodynamics is associated with thrombotic risk after fenestrated endovascular aneurysm repair with large diameter endografts.

Authors:  Kenneth Tran; K Brennan Feliciano; Weiguang Yang; Erica L Schwarz; Alison L Marsden; Ronald L Dalman; Jason T Lee
Journal:  JVS Vasc Sci       Date:  2022-04-21

2.  FDA's nozzle numerical simulation challenge: non-Newtonian fluid effects and blood damage.

Authors:  Miquel Trias; Antonio Arbona; Joan Massó; Borja Miñano; Carles Bona
Journal:  PLoS One       Date:  2014-03-25       Impact factor: 3.240

3.  Patient-specific computational flow modelling for assessing hemodynamic changes following fenestrated endovascular aneurysm repair.

Authors:  Kenneth Tran; Weiguang Yang; Alison Marsden; Jason T Lee
Journal:  JVS Vasc Sci       Date:  2021-03-03
  3 in total