A novel energy mapping approach for CT-based attenuation correction in PET.

B Teimourian1, M R Ay, M Shamsaie Zafarghandi, P Ghafarian, H Ghadiri, H Zaidi.   

Abstract

PURPOSE: Dual-energy CT (DECT) is arguably the most accurate energy mapping technique in CT-based attenuation correction (CTAC) implemented on hybrid PET/CT systems. However, this approach is not attractive for clinical use owing to increased patient dose. The authors propose a novel energy mapping approach referred to as virtual DECT (VDECT) taking advantage of the DECT formulation but using CT data acquired at a single energy (kV(P)). For this purpose, the CT image acquired at one energy is used to generate the CT image at a second energy using calculated kV(P) conversion curves derived from phantom studies.
METHODS: The attenuation map (μ-map) at 511 keV was generated for the XCAT phantom and clinical studies using the bilinear, DECT, and VDECT techniques. The generated μ-maps at 511 keV are compared to the reference derived from the XCAT phantom serving as ground truth. PET data generated from a predefined activity map for the XCAT phantom were then corrected for attenuation using μ-maps generated using the different energy mapping approaches. In addition, the generated μ-maps using the above described methods for a cylindrical polyethylene phantom containing different concentrations of K(2)HPO(4) in water were compared to actual attenuation coefficients. Likewise, CT images of five clinical whole-body studies were used to generate μ-maps using the various energy-mapping approaches were compared with μ-maps acquired at 511 keV using (68)Ge/(68)Ga rod sources for the clinical studies.
RESULTS: The results of phantom studies demonstrate that the proposed method is more accurate than the bilinear technique. All three μ-maps yielded almost similar results for soft and lung tissues whereas for bone tissues, the DECT and the VDECT methods produced a much smaller mean relative difference (3.0% and 2.8%, respectively) than the bilinear approach (11.8%). Likewise, the comparison of PET images corrected for attenuation using the various methods showed that the proposed method provides better accuracy (6.5%) than the bilinear method (13.4%). Clinical studies further demonstrated that, compared to the bilinear method, the VDECT approach has better agreement for bony structures with the DECT technique (1.5% versus 8.9%) and transmission scanning (8.8% versus 17.7%).
CONCLUSIONS: It was concluded that the proposed method outperforms the bilinear method especially in bony structures. Further evaluation using a large clinical PET/CT database is underway to evaluate the potential of the technique in a clinical setting.

Entities:  

Mesh:

Year:  2012        PMID: 22482629     DOI: 10.1118/1.3694108

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Dual energy CT for attenuation correction with PET/CT.

Authors:  Ting Xia; Adam M Alessio; Paul E Kinahan
Journal:  Med Phys       Date:  2014-01       Impact factor: 4.071

2.  A New Approach for Scatter Removal and Attenuation Compensation from SPECT/CT Images.

Authors:  Shabnam Oloomi; Hadi Noori Eskandari; Seyed Rasoul Zakavi; Peter Knoll; Faraz Kalantari; Mohsen Hajizadeh Saffar
Journal:  Iran J Basic Med Sci       Date:  2013-11       Impact factor: 2.699

Review 3.  PET/MRI attenuation estimation in the lung: A review of past, present, and potential techniques.

Authors:  Joseph Lillington; Ludovica Brusaferri; Kerstin Kläser; Karin Shmueli; Radhouene Neji; Brian F Hutton; Francesco Fraioli; Simon Arridge; Manuel Jorge Cardoso; Sebastien Ourselin; Kris Thielemans; David Atkinson
Journal:  Med Phys       Date:  2020-01-01       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.