Characteristics of miniature electronic brachytherapy x-ray sources based on TG-43U1 formalism using Monte Carlo simulation techniques.

Habib Safigholi1, Reza Faghihi, Somaye Karimi Jashni, Ali S Meigooni.   

Abstract

PURPOSE: The goal of this study is to determine a method for Monte Carlo (MC) characterization of the miniature electronic brachytherapy x-ray sources (MEBXS) and to set dosimetric parameters according to TG-43U1 formalism. TG-43U1 parameters were used to get optimal designs of MEBXS. Parameters that affect the dose distribution such as anode shapes, target thickness, target angles, and electron beam source characteristics were evaluated. Optimized MEBXS designs were obtained and used to determine radial dose functions and 2D anisotropy functions in the electron energy range of 25-80 keV.
METHODS: Tungsten anode material was considered in two different geometries, hemispherical and conical-hemisphere. These configurations were analyzed by the 4C MC code with several different optimization techniques. The first optimization compared target thickness layers versus electron energy. These optimized thicknesses were compared with published results by Ihsan et al. [Nucl. Instrum. Methods Phys. Res. B 264, 371-377 (2007)]. The second optimization evaluated electron source characteristics by changing the cathode shapes and electron energies. Electron sources studied included; (1) point sources, (2) uniform cylinders, and (3) nonuniform cylindrical shell geometries. The third optimization was used to assess the apex angle of the conical-hemisphere target. The goal of these optimizations was to produce 2D-dose anisotropy functions closer to unity. An overall optimized MEBXS was developed from this analysis. The results obtained from this model were compared to known characteristics of HDR (125)I, LDR (103)Pd, and Xoft Axxent™ electronic brachytherapy source (XAEBS) [Med. Phys. 33, 4020-4032 (2006)].
RESULTS: The optimized anode thicknesses as a function of electron energy is fitted by the linear equation Y (μm) = 0.0459X (keV)-0.7342. The optimized electron source geometry is obtained for a disk-shaped parallel beam (uniform cylinder) with 0.9 mm radius. The TG-43 distribution is less sensitive to the shape of the conical-hemisphere anode than the hemispherical anode. However, the optimized apex angle of conical-hemisphere anode was determined to be 60°. For the hemispherical targets, calculated radial dose function values at a distance of 5 cm were 0.137, 0.191, 0.247, and 0.331 for 40, 50, 60, and 80 keV electrons, respectively. These values for the conical-hemisphere targets are 0.165, 0.239, 0.305, and 0.412, respectively. Calculated 2D anisotropy functions values for the hemispherical target shape were F(1 cm, 0°) = 1.438 and F(1 cm, 0°) = 1.465 for 30 and 80 keV electrons, respectively. The corresponding values for conical-hemisphere targets are 1.091 and 1.241, respectively.
CONCLUSIONS: A method for the characterizations of MEBXS using TG-43U1 dosimetric data using the MC MCNP4C has been presented. The effects of target geometry, thicknesses, and electron source geometry have been investigated. The final choices of MEBXS design are conical-hemisphere target shapes having an apex angle of 60°. Tungsten material having an optimized thickness versus electron energy and a 0.9 mm radius of uniform cylinder as a cathode produces optimal electron source characteristics.

Entities:  

Mesh:

Year:  2012        PMID: 22482618     DOI: 10.1118/1.3693046

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

Review 1.  Electronic brachytherapy--current status and future directions.

Authors:  D J Eaton
Journal:  Br J Radiol       Date:  2015-03-06       Impact factor: 3.039

2.  Anode optimization for miniature electronic brachytherapy X-ray sources using Monte Carlo and computational fluid dynamic codes.

Authors:  Masoud Khajeh; Habib Safigholi
Journal:  J Adv Res       Date:  2015-04-20       Impact factor: 10.479

3.  Assessment of two hemispherical and hemispherical-conical miniature sources used in electronic brachytherapy using Monte Carlo Simulation.

Authors:  Barat Barati; Mansour Zabihzadeh; Mohammad Javad Tahmasebi Birgani; Nahid Chegini; Mojtaba Hoseini Ghahfarokhi; Jafar Fatahiasl
Journal:  Electron Physician       Date:  2017-02-25

4.  Optimum radiation source for radiation therapy of skin cancer.

Authors:  Habib Safigholi; William Y Song; Ali S Meigooni
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

5.  Treatment of cervical cancer with electronic brachytherapy.

Authors:  Sergio Lozares-Cordero; José Antonio Font-Gómez; Almudena Gandía-Martínez; Anabela Miranda-Burgos; Agustina Méndez-Villamón; David Villa-Gazulla; Verónica Alba-Escorihuela; Sara Jiménez-Puertas; Víctor González-Pérez
Journal:  J Appl Clin Med Phys       Date:  2019-06-11       Impact factor: 2.102