In vivo dosimetry for gynaecological brachytherapy using a novel position sensitive radiation detector: feasibility study.

B Reniers1, G Landry, R Eichner, A Hallil, F Verhaegen.   

Abstract

PURPOSE: In gynecological radiotherapy with high dose rate (HDR)(192)Ir brachytherapy, the treatment complexity has increased due to improved optimization techniques and dose constraints. As a consequence, it has become more important to verify the dose delivery to the target and also to the organs at risk (e.g., the bladder). In vivo dosimetry, where dosimeters are placed in or on the patient, is one way of verifying the dose but until recently this was hampered by motion of the radiation detectors with respect to the source. The authors present a novel dosimetry method using a position sensitive radiation detector.
METHODS: The prototype RADPOS system (Best Medical Canada) consists of a metal oxide field effect transistor (MOSFET) dosimeter coupled to a position-sensor, which deduces its 3D position in a magnetic field. To assess the feasibility of in vivo dosimetry based on the RADPOS system, different characteristics of the detector need to be investigated. Using a PMMA phantom, the positioning accuracy of the RADPOS system was quantified by comparing position readouts with the known position of the detector along the x and y-axes. RADPOS dose measurements were performed at various distances from a Nucletron(192)Ir source in a PMMA phantom to evaluate the energy dependence of the MOSFET. A sensitivity analysis was performed by calculating the dose after varying (1) the position of the RADPOS detector to simulate organ motion and (2) the position of the first dwell position to simulate errors in delivery. The authors also performed an uncertainty analysis to determine the action level (AL) that should be used during in vivo dosimetry.
RESULTS: Positioning accuracy is found to be within 1 mm in the 1-10 cm range from the origin along the x-axis (away from the transmitter), meeting the requirements for in vivo dosimetry. Similar results are obtained for the other axes. The ALs are chosen to take into account the total uncertainty on the measurements. As a consequence for in vivo dosimetry, it is determined that the RADPOS sensor, if placed, for example, in the bladder Foley balloon, would detect a 2 mm motion of the bladder, at a 5% chance of a false positive, with an AL limit of 9% of the dose delivered. The authors found that source position errors, caused by, e.g., a wrong first dwell position, are more difficult to detect; indeed, with our single RADPOS detector, positioned in the bladder, dwell position errors below 5 mm and resulting in a dose error within 10%, could be detected in the tandem but not in the colpostats. A possible solution to improve error detection is to use multiple MOSFETs to obtain multiple dose values.
CONCLUSIONS: In this study, the authors proposed a dosimetry procedure, based on the novel RADPOS system, to accurately determine the position of the radiation dosimeter with respect to the applicator. The authors found that it is possible to monitor the delivered dose in a point and compare it to the predetermined dose. This allows in principle the detection of problems such as bladder motion/filling or source mispositioning. Further clinical investigation is warranted.

Entities:  

Mesh:

Year:  2012        PMID: 22482614     DOI: 10.1118/1.3693049

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  On the use of a single-fiber multipoint plastic scintillation detector for 192Ir high-dose-rate brachytherapy.

Authors:  François Therriault-Proulx; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

2.  An investigation of a PRESAGE® in vivo dosimeter for brachytherapy.

Authors:  A K Vidovic; T Juang; S Meltsner; J Adamovics; J Chino; B Steffey; O Craciunescu; M Oldham
Journal:  Phys Med Biol       Date:  2014-06-24       Impact factor: 3.609

3.  Inorganic scintillation detectors based on Eu-activated phosphors for 192Ir brachytherapy.

Authors:  Gustavo Kertzscher; Sam Beddar
Journal:  Phys Med Biol       Date:  2017-05-05       Impact factor: 3.609

Review 4.  In vivo dosimetry: trends and prospects for brachytherapy.

Authors:  G Kertzscher; A Rosenfeld; S Beddar; K Tanderup; J E Cygler
Journal:  Br J Radiol       Date:  2014-07-08       Impact factor: 3.039

5.  Ruby-based inorganic scintillation detectors for 192Ir brachytherapy.

Authors:  Gustavo Kertzscher; Sam Beddar
Journal:  Phys Med Biol       Date:  2016-10-14       Impact factor: 3.609

Review 6.  In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Gabriel P Fonseca; Jacob G Johansen; Ryan L Smith; Luc Beaulieu; Sam Beddar; Gustavo Kertzscher; Frank Verhaegen; Kari Tanderup
Journal:  Phys Imaging Radiat Oncol       Date:  2020-09-28

7.  3D in vivo dosimetry of HDR gynecological brachytherapy using micro silica bead TLDs.

Authors:  Ramin Jaberi; Somayyeh Babaloui; Zahra Siavashpour; Maryam Moshtaghi; Alireza Shirazi; Musa Joya; Mohammad Hadi Gholami; Shakardokht Jafari
Journal:  J Appl Clin Med Phys       Date:  2022-08-10       Impact factor: 2.243

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.