Validation of GPU based TomoTherapy dose calculation engine.

Quan Chen1, Weiguo Lu, Yu Chen, Mingli Chen, Douglas Henderson, Edmond Sterpin.   

Abstract

PURPOSE: The graphic processing unit (GPU) based TomoTherapy convolution/superposition(C/S) dose engine (GPU dose engine) achieves a dramatic performance improvement over the traditional CPU-cluster based TomoTherapy dose engine (CPU dose engine). Besides the architecture difference between the GPU and CPU, there are several algorithm changes from the CPU dose engine to the GPU dose engine. These changes made the GPU dose slightly different from the CPU-cluster dose. In order for the commercial release of the GPU dose engine, its accuracy has to be validated.
METHODS: Thirty eight TomoTherapy phantom plans and 19 patient plans were calculated with both dose engines to evaluate the equivalency between the two dose engines. Gamma indices (Γ) were used for the equivalency evaluation. The GPU dose was further verified with the absolute point dose measurement with ion chamber and film measurements for phantom plans. Monte Carlo calculation was used as a reference for both dose engines in the accuracy evaluation in heterogeneous phantom and actual patients.
RESULTS: The GPU dose engine showed excellent agreement with the current CPU dose engine. The majority of cases had over 99.99% of voxels with Γ(1%, 1 mm) < 1. The worst case observed in the phantom had 0.22% voxels violating the criterion. In patient cases, the worst percentage of voxels violating the criterion was 0.57%. For absolute point dose verification, all cases agreed with measurement to within ±3% with average error magnitude within 1%. All cases passed the acceptance criterion that more than 95% of the pixels have Γ(3%, 3 mm) < 1 in film measurement, and the average passing pixel percentage is 98.5%-99%. The GPU dose engine also showed similar degree of accuracy in heterogeneous media as the current TomoTherapy dose engine.
CONCLUSIONS: It is verified and validated that the ultrafast TomoTherapy GPU dose engine can safely replace the existing TomoTherapy cluster based dose engine without degradation in dose accuracy.

Entities:  

Mesh:

Year:  2012        PMID: 22482609     DOI: 10.1118/1.3693057

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Parallel beamlet dose calculation via beamlet contexts in a distributed multi-GPU framework.

Authors:  Ryan Neph; Cheng Ouyang; John Neylon; Youming Yang; Ke Sheng
Journal:  Med Phys       Date:  2019-06-30       Impact factor: 4.071

2.  Dosimetric evaluation of intensity-modulated radiotherapy, volumetric modulated arc therapy, and helical tomotherapy for hippocampal-avoidance whole brain radiotherapy.

Authors:  Yi Rong; Josh Evans; Meng Xu-Welliver; Cadron Pickett; Guang Jia; Quan Chen; Li Zuo
Journal:  PLoS One       Date:  2015-04-20       Impact factor: 3.240

3.  Parallel/Opposed: IMRT QA using treatment log files is superior to conventional measurement-based method.

Authors:  Nathan Childress; Quan Chen; Yi Rong
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

4.  A virtual source model for Monte Carlo simulation of helical tomotherapy.

Authors:  Jiankui Yuan; Yi Rong; Quan Chen
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

5.  Radiation dose escalation based on FDG-PET driven dose painting by numbers in oropharyngeal squamous cell carcinoma: a dosimetric comparison between TomoTherapy-HA and RapidArc.

Authors:  Sarah Differding; Edmond Sterpin; Nicolas Hermand; Bianca Vanstraelen; Sandra Nuyts; Nathalie de Patoul; Jean-Marc Denis; John Aldo Lee; Vincent Grégoire
Journal:  Radiat Oncol       Date:  2017-03-23       Impact factor: 3.481

6.  Analysis of clinical patient-specific pre-treatment quality assurance with the new helical tomotherapy platform, following the AAPM TG-218 report.

Authors:  Marco Fusella; Samuele Cavinato; Alessandra Germani; Marta Paiusco; Nicola Pivato; Marco Andrea Rossato; Anthony Scott; Alessandro Scaggion
Journal:  Radiat Oncol       Date:  2021-11-22       Impact factor: 3.481

7.  Performance Characteristics of an Independent Dose Verification Program for Helical Tomotherapy.

Authors:  Isaac C F Chang; Jeff Chen; Slav Yartsev
Journal:  J Med Phys       Date:  2017 Jul-Sep
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.