How to improve x-ray scattering techniques to quantify bone mineral density using spectroscopy.

M Krmar1, K Ganezer.   

Abstract

PURPOSE: The purpose of this study was to develop a new diagnostic technique for measuring bone mineral density (BMD) for the assessment of osteoporosis, which improves upon the coherent to Compton scattering ratio (CCSR) method, which was first developed in the 1980s. To help the authors achieve these goals, they have identified and studied two new indices for CCSR, the forward scattered to backward scattered (FS-BS) and the forward scattered to transmitted (FS-T) ratios. They believe that, at small angles, these two parameters can offer a practical in vivo determination of BMD that can be used to overcome the limitations of past CCSR systems, including high radiation dosages, costs, and examination durations.
METHODS: In previous CCSR studies, a high-activity radioactive source with a long half-live (usually (241)Am) and an expensive and bulky cryogenic HPGe detector were applied to both in vivo and in vitro measurements. To make this technique more suitable for clinical applications, the possibility of using a standard diagnostic x-ray tube generating a continuous spectrum was investigated in this paper. Scattered radiation from trabecular bone-simulating phantoms containing various mineral densities that span the normal range of in vivo BMD was collected in this study using relatively inexpensive noncryogenic CdTe or NaI detectors.
RESULTS: The initial results demonstrate that a modified version of CCSR can be successfully applied to trabecular bone assessment using a diagnostic x-ray tube with a continuous spectrum in two variations, the FS-BS and the FS-T ratio. When FS-BS is measured, intensity spectra in the forward and backward directions must be collected while FS-T requires only the integral intensity of the scattered and transmitted (T) spectra in the energy region above 40 keV. For both of these methods, forward scattering angles less than or equal to 15° and backward scattering angles greater than or equal to (165°= 180° - 15°) are needed.
CONCLUSIONS: The authors determined that FS-T is more sensitive to changes in BMD than transmission or absorption alone and that the FS-BS method can yield an absolute measurement of the mean atomic number of the scattering medium, after a correction for path-dependent attenuation. Since this study determined that the FS-T ratio is independent of the incident energy over a broad energy region, it will be possible to apply FS-T to bone densitometry using inexpensive integral photon detectors. The authors believe that, by replacing the radionuclide source with an x-ray tube and the cryogenically cooled HPGe detector with a single solid state CdTe, NaI, or silicon detector or an annular array of detectors, as suggested in this study, the past difficulties of CCSR concerning high radiation exposure, costs, and durations as well as lack of convenience can be overcome and that CCSR could eventually become popular in clinical settings.

Entities:  

Mesh:

Year:  2012        PMID: 22482605      PMCID: PMC3316692          DOI: 10.1118/1.3687162

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

Review 1.  Current status of physical measurements of the skeleton.

Authors:  M A Greenfield
Journal:  Med Phys       Date:  1992 Nov-Dec       Impact factor: 4.071

2.  Measurements of electron densities with the aid of the Compton scattering process.

Authors:  E ODEBLAD; A NORHAGEN
Journal:  Acta radiol       Date:  1956-02       Impact factor: 1.990

Review 3.  Methods of bone mineral measurement.

Authors:  P Tothill
Journal:  Phys Med Biol       Date:  1989-05       Impact factor: 3.609

Review 4.  Noninvasive assessment of bone mineral and structure: state of the art.

Authors:  H K Genant; K Engelke; T Fuerst; C C Glüer; S Grampp; S T Harris; M Jergas; T Lang; Y Lu; S Majumdar; A Mathur; M Takada
Journal:  J Bone Miner Res       Date:  1996-06       Impact factor: 6.741

5.  A method for simultaneous correction of spectrum hardening artifacts in CT images containing both bone and iodine.

Authors:  P M Joseph; C Ruth
Journal:  Med Phys       Date:  1997-10       Impact factor: 4.071

6.  The effect of the momentum transfer on the sensitivity of a photon scattering method for the characterization of tissues.

Authors:  I Leichter; A Karellas; J D Craven; M A Greenfield
Journal:  Med Phys       Date:  1984 Jan-Feb       Impact factor: 4.071

7.  A large-angle coherent/Compton scattering method for measurement in vitro of trabecular bone mineral concentration.

Authors:  G E Gigante; S Sciuti
Journal:  Med Phys       Date:  1985 May-Jun       Impact factor: 4.071

8.  The measurement of trabecular bone mineral density using coherent and Compton scattered photons in vitro.

Authors:  S S Ling; S Rustgi; A Karellas; J D Craven; J S Whiting; M A Greenfield; R Stern
Journal:  Med Phys       Date:  1982 Mar-Apr       Impact factor: 4.071

9.  Quantitative assessment of bone mineral by photon scattering: calibration considerations.

Authors:  I Leichter; A Karellas; S S Shukla; J L Looper; J D Craven; M A Greenfield
Journal:  Med Phys       Date:  1985 Jul-Aug       Impact factor: 4.071

10.  Trabecular bone mineral density measurement in vivo: use of the ratio of coherent to Compton-scattered photons in the calcaneus.

Authors:  S S Shukla; I Leichter; A Karellas; J D Craven; M A Greenfield
Journal:  Radiology       Date:  1986-03       Impact factor: 11.105