Literature DB >> 22467152

Safety and efficacy of StarClose SE Vascular Closure System in high-risk liver interventional oncology patients.

Ron C Gaba1, Ahmad Parvinian, Estrellita M Trinos, Suzanne V Padayao, Rechel M Francisco, Felix Y Yap, M Grace Knuttinen, Charles A Owens, James T Bui.   

Abstract

PURPOSE: To assess the safety and efficacy of the StarClose SE Vascular Closure System (Abbott Vascular, Abbott Park IL, USA) in high-risk thrombocytopenic and coagulopathic interventional oncology (IO) patients.
METHODS: In this single institution retrospective study, 63 high-risk thrombocytopenic or coagulopathic IO patients (M:F=51:12, mean age 58 years, range 31-88 years) who underwent 83 common femoral arteriotomy closures using the StarClose device were identified among all IO patients (n=131) undergoing StarClose closure (n=177) between 2008-2011. High-risk thrombocytopenia and coagulopathy were defined as platelet count ≤100 10(3)/mL and international normalized ratio (INR) ≥1.5. Procedures included chemoembolization (n=67), radioembolization (n=8), and hepatic arterial mapping with technetium-99m macroaggrated albumin administration (n=8) for treatment of hepatocellular carcinoma (n=79) or liver metastases (n=4). Measured outcomes included technical success of arterial closure and closure-related adverse events, graded according to the Society of Interventional Radiology classification.
RESULTS: In all cases, 5 French common femoral arterial access was used. Platelet count was ≤100 10(3)/mL in 80/83 (96.4%) cases and INR was ≥1.5 in 35/83 (42.2%) cases. Mean pre-procedure platelet count was 71 (range 26-347) 10(3)/mL and mean INR was 1.4 (range 1.0-2.1). The StarClose device effectively sealed the arteriotomy in 83/83 (100%) cases, 60/83 (72.3%) cases were first-time closures, and 20/83 (24.1%) cases were repeat closures. Small groin hematomas, graded as class A minor complications, developed in 3/83 (3.6%) cases. No other complications were encountered.
CONCLUSIONS: The StarClose SE Vascular Closure System confers high technical success and safety in common femoral arteriotomy closure in high-risk IO patients.

Entities:  

Mesh:

Year:  2012        PMID: 22467152     DOI: 10.5301/jva.5000068

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  1 in total

1.  Flip-over of blood vessel intima caused by vascular closure device: A case report.

Authors:  Li-Xia Sun; Xue-Song Yang; Da-Wei Zhang; Bo Zhao; Lian-Lian Li; Qi Zhang; Qing-Zhi Hao
Journal:  World J Clin Cases       Date:  2022-03-06       Impact factor: 1.337

  1 in total