Transcutaneous bilirubin measurement: comparison of Respironics BiliCheck and JM-103 in a normal newborn population.

Costantino Romagnoli1, Enrico Zecca, Piero Catenazzi, Giovanni Barone, Antonio Alberto Zuppa.   

Abstract

OBJECTIVES: To compare the accuracy of BiliCheck™ (Respironics, Marietta, GA) and Konica-Minolta Air Shield JM-103 (Drager Medical Inc, Telford, PA) to evaluate total serum bilirubin (TSB).
METHODS: Prospective blinded study comparing two diagnostic devices in 630 neonates requiring TSB measurement.
RESULTS: Linear regression analysis showed a good correlation between BiliCheck™ and TSB (r=0.8212) as well as between JM-103 and TSB (r=0.8686). BiliCheck shows a tendency to underestimate TSB. The mean difference in TSB-TcB was -1.4 mg/dL for BC (-4.7/+1.8 mg/dL) and 0.3 mg/dL for JM-103 (-2.6/+3.2mg/dL). ROC analysis for TSB≥ 12 mg/dL showed area under the curve for BiliCheck™ significantly lower than those for JM-103 (p<0.0001). JM-103 resulted less time expensive than BiliCheck.
CONCLUSIONS: In spite of similar diagnostic accuracy JM-103 could be preferred for some practical advantages, but its suitability in performing universal screening for severe hyperbilirubinemia deserves further investigations.
Copyright © 2012 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22465272     DOI: 10.1016/j.clinbiochem.2012.03.014

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  8 in total

1.  Screening neonatal jaundice based on the sclera color of the eye using digital photography.

Authors:  Terence S Leung; Karan Kapur; Ashley Guilliam; Jade Okell; Bee Lim; Lindsay W MacDonald; Judith Meek
Journal:  Biomed Opt Express       Date:  2015-10-23       Impact factor: 3.732

Review 2.  Bilirubin nomograms for identification of neonatal hyperbilirubinemia in healthy term and late-preterm infants: a systematic review and meta-analysis.

Authors:  Zhang-Bin Yu; Shu-Ping Han; Chao Chen
Journal:  World J Pediatr       Date:  2014-08-15       Impact factor: 2.764

3.  BiliCheck vs JM-103 in identifying neonates not at risk of hyperbilirubinaemia.

Authors:  Costantino Romagnoli; Piero Catenazzi; Giovanni Barone; Lucia Giordano; Riccardo Riccardi; Antonio Alberto Zuppa; Enrico Zecca
Journal:  Ital J Pediatr       Date:  2013-07-23       Impact factor: 2.638

4.  Italian guidelines for management and treatment of hyperbilirubinaemia of newborn infants ≥ 35 weeks' gestational age.

Authors:  Costantino Romagnoli; Giovanni Barone; Simone Pratesi; Francesco Raimondi; Letizia Capasso; Enrico Zecca; Carlo Dani
Journal:  Ital J Pediatr       Date:  2014-01-31       Impact factor: 2.638

5.  Transcutaneous bilirubin nomograms in African neonates.

Authors:  Bolajoko O Olusanya; Cecilia A Mabogunje; Donald O Imosemi; Abieyuwa A Emokpae
Journal:  PLoS One       Date:  2017-02-13       Impact factor: 3.240

6.  Screening and treatment to reduce severe hyperbilirubinaemia in infants in primary care (STARSHIP): a factorial stepped-wedge cluster randomised controlled trial protocol.

Authors:  Berthe A M van der Geest; Johanna P de Graaf; Loes C M Bertens; Marten J Poley; Erwin Ista; René F Kornelisse; Irwin K M Reiss; Eric A P Steegers; Jasper V Been
Journal:  BMJ Open       Date:  2019-04-20       Impact factor: 2.692

7.  Transcutaneous bilirubin measurement in healthy Saudi term newborns.

Authors:  Saad A Alsaedi
Journal:  Saudi Med J       Date:  2016-02       Impact factor: 1.484

8.  Transcutaneous Bilirubin Measurements Can Be Used to Measure Bilirubin Levels during Phototherapy.

Authors:  Saad Abdullah Alsaedi
Journal:  Int J Pediatr       Date:  2018-03-20
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.