Retrieval analysis of squeaking ceramic implants: are there related specific features?

C Chevillotte1, R T Trousdale, K-N An, D Padgett, T Wright.   

Abstract

INTRODUCTION: Ceramic-on-ceramic total hip arthroplasty is routinely used for young and active patients with end stage of hip osteoarthritis. However, squeaking noise is a recently identified problem with such bearing surface. Many in vivo and in vitro studies have been conducted trying to find the potential causes of this phenomenon. However, we are not aware of any study analyzing retrieved ceramic implants for squeaking. HYPOTHESIS: Our primary hypothesis was that the surface analysis of retrieved ceramic implants with squeaking would present interesting deteriorations that could explain the squeaking noise.
MATERIALS AND METHODS: Nine retrieved squeaking implants from ceramic-on-ceramic total hip arthroplasty that were retrieved for various reasons (two exclusively for squeaking, four for recurrent dislocation, one for aseptic loosening and two for instability) were analyzed. Implant positioning was calculated, macroscopic damages were noticed and microscopic roughness was analyzed. The retrieved implants were then tested on a hip simulator reproducing flexion/extension motions in several situations in lubricated and non-lubricated conditions in order to reproduce squeaking.
RESULTS: Five cups were considered with borderline insufficient anteversion. Gross impingement damage was visible on seven implants. All the retrieved heads had visible metal transfer on their surface. Eight implants had visible stripe wear. Microscopic analysis showed roughness higher than six microns on the retrieved heads. Squeaking was reproduced in vitro in dry conditions. In lubricated conditions, squeaking did not occur for the retrieved hips. DISCUSSION: This retrieval analysis suggests that problems of cup orientation and design which can lead to impingement can generate lubrication problems because of metal transfer plus/minus stripe wear which is a common theme in ceramic-on-ceramic bearings that squeak.
Copyright © 2012 Elsevier Masson SAS. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22459099     DOI: 10.1016/j.otsr.2011.12.003

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  10 in total

1.  Difference in the acetabular cup orientation in standing and supine radiographs.

Authors:  Munir Khan; Tom Beckingsale; Martin Marsh; Jim Holland
Journal:  J Orthop       Date:  2016-04-05

2.  Meta-analysis of the Incidence and Risk Factors for Squeaking after Primary Ceramic-on-ceramic Total Hip Arthroplasty in Asian Patients.

Authors:  Tae-Hun Lee; Young-Wan Moon; Seung-Jae Lim; Youn-Soo Park
Journal:  Hip Pelvis       Date:  2014-06-30

3.  Сomparative analysis of large diameter bearing ceramic monoblock acetabular components.

Authors:  Valery Yu Murylev; Pavel M Elizarov; Aleksei V Muzychenkov; Yaroslav A Rukin; Grigory A Kukovenko; Semen S Alekseev; Michail P Elizarov
Journal:  J Orthop       Date:  2022-08-11

4.  High failure rate of a new pressfit cup in mid-term follow-up.

Authors:  Steffen Brodt; Georg Matziolis; Christoph Windisch; Andreas Gosse; Matthias Spalteholz; Ralf Herbert Gahr
Journal:  Int Orthop       Date:  2015-07-08       Impact factor: 3.075

5.  Do the Reasons for Ceramic-on-ceramic Revisions Differ From Other Bearings in Total Hip Arthroplasty?

Authors:  Henri Migaud; Sophie Putman; Grégory Kern; Ronald Isida; Julien Girard; Nassima Ramdane; Christian P Delaunay; Moussa Hamadouche
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

6.  Ceramic-on-ceramic THA Implants in Patients Younger Than 20 Years.

Authors:  Didier Hannouche; Flore Devriese; Jérôme Delambre; Frédéric Zadegan; Idriss Tourabaly; Laurent Sedel; Sylvie Chevret; Rémy Nizard
Journal:  Clin Orthop Relat Res       Date:  2015-09-04       Impact factor: 4.176

7.  Tribology of total hip arthroplasty prostheses: What an orthopaedic surgeon should know.

Authors:  Claude B Rieker
Journal:  EFORT Open Rev       Date:  2017-03-13

8.  Third-generation pure alumina and alumina matrix composites in total hip arthroplasty: What is the evidence?

Authors:  Didier Hannouche; Matthieu Zingg; Hermes Miozzari; Remy Nizard; Anne Lübbeke
Journal:  EFORT Open Rev       Date:  2018-01-10

9.  Arthroscopic evaluation and treatment of a squeaking hip. A case report.

Authors:  Jonathan Bellity; Marc Elkaïm; Didier Hannouche; Rémy Nizard
Journal:  BMC Musculoskelet Disord       Date:  2020-12-03       Impact factor: 2.362

Review 10.  Hip Squeaking after Ceramic-on-ceramic Total Hip Arthroplasty.

Authors:  Guo-Liang Wu; Wei Zhu; Yan Zhao; Qi Ma; Xi-Sheng Weng
Journal:  Chin Med J (Engl)       Date:  2016-08-05       Impact factor: 2.628

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.