Literature DB >> 22458866

Comparison of conventional and digital radiography systems with regard to radiopacity of root canal filling materials.

I Akcay1, B Ilhan, N Dundar.   

Abstract

AIM: To compare the radiopacity of five root filling materials as measured on digitized E-speed films, storage phosphor plate and CCD sensor images.
METHODOLOGY: Radiopacity of Resilon cones and Epiphany, MM-Seal, EndoREZ and AH Plus sealers were investigated. Ten acrylic plates were prepared and filled with the materials. Samples were radiographed together with samples of dentine and an aluminium stepwedge. Images were obtained using E-speed films, Digora storage phosphor plates and Schick CDR CCD sensor. Conventional radiographs were digitized using a desktop scanner. Mean grey values of the materials, stepwedge and dentine were measured using the histogram analysis function of Photoshop 7.0. A graph of radiographic density versus thickness of the aluminium was drawn for each image from which the radiographic densities of the materials were determined in relation to the aluminium thickness. The mean was calculated, and the data were analysed using repeated measures anova and Bonferroni tests (P<0.05).
RESULTS: There were significant differences between different imaging modalities (P=0.018) and root filling materials (P<0.001). A statistically significant interaction between imaging modality and material was also observed (P<0.001). All materials had radiopacity above 3mm of aluminium regardless of the imaging technique. The radiopacity of Resilon and Epiphany was significantly higher on digital images compared with digitized E-speed film images (P<0.001).
CONCLUSION: The choice of imaging system may effect radiopacity measurements. It is possible that radiopacity as recorded on traditional or digitized films is not indicative of the radiopacity as recorded on a digital sensor.
© 2012 International Endodontic Journal.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22458866     DOI: 10.1111/j.1365-2591.2012.02026.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  5 in total

1.  A new in vitro method to evaluate radio-opacity of endodontic sealers.

Authors:  V B Malka; G L Hochscheidt; N L Larentis; F S Grecca; V R C Fontanella; P M P Kopper
Journal:  Dentomaxillofac Radiol       Date:  2015-02-20       Impact factor: 2.419

2.  [Comparative study on radio-opacity of dental composite resin materials'determination using film imaging and digital imaging].

Authors:  Y Li; H Lin; T J Zhang
Journal:  Beijing Da Xue Xue Bao Yi Xue Ban       Date:  2021-10-18

Review 3.  Mineral Trioxide Aggregate-A Review of Properties and Testing Methodologies.

Authors:  William N Ha; Timothy Nicholson; Bill Kahler; Laurence J Walsh
Journal:  Materials (Basel)       Date:  2017-11-02       Impact factor: 3.623

4.  The best radiographic method for determining root canal morphology in mandibular first premolars: A study of Chinese descendants in Taiwan.

Authors:  Yu Sun; Tzu-Yi Lu; Yi-Chen Chen; Shue-Fen Yang
Journal:  J Dent Sci       Date:  2016-04-12       Impact factor: 2.080

5.  Radiopacity evaluation of contemporary resin composites by digitization of images.

Authors:  R Banu Ermis; Derya Yildirim; Gul Yildiz; Ozlem Gormez
Journal:  Eur J Dent       Date:  2014-07