Heart perforation in patients with permanent cardiac pacing - pilot personal observations.

Justyna Piekarz1, Jacek Lelakowski, Anna Rydlewska, Jacek Majewski.   

Abstract

INTRODUCTION: Heart perforation is a rare complication of pacemaker (PM)/implantable cardioverter-defibrillator (ICD) implantation.
MATERIAL AND METHODS: In our clinic in 2005-2010, 6 patients with heart perforation were hospitalized (3 women, 3 men), mean age 58.6 ±20.8 years (17 to 73 years). The indication to PM/ICD implantation was tachy-brady syndrome in 3 cases, second-degree atrioventricular block, advanced with losses of consciousness, vaso-vagal syndrome type II B with asystole lasting 12 s and recurrent non-sustained ventricular tachycardia in 1 patient. We analyzed patient's medical records, X-rays, echocardiography, computed tomography (CT) and procedure protocols.
RESULTS: The incidence of heart perforation was 0.09%. Symptoms developed 4 to 990 days (mean 186.3 ±394.3) after PM/ICD implantation. The perforation site was found in the right atrial wall in 1 cases and the right ventricular wall in 6 cases. The TTE revealed an accumulation of fluid in the pericardium over 10 mm behind the posterior wall of the left ventricle in all patients. The CT scan confirmed perforation of the heart chambers (atrium and in 6 cases ventricle). In 5 cases the whole device was removed by direct traction or percutaneous lead extraction with pericardiocentesis when necessary (pericardium drainage in 3 cases) while in 1 case cardiac surgery was needed.
CONCLUSIONS: The perforating lead may be removed by direct traction in the operating room with cardiosurgical, anesthesiological and echocardiographical backup. In case of the lead perforation outside the pericardial sac or its atypical location, cardiac surgery is a safer method. The most important diagnostic method remains computed tomography.

Entities:  

Keywords:  heart perforation; pacemaker/cardioverter-defibrillator implantation

Year:  2012        PMID: 22457678      PMCID: PMC3309440          DOI: 10.5114/aoms.2012.27284

Source DB:  PubMed          Journal:  Arch Med Sci        ISSN: 1734-1922            Impact factor:   3.318


Introduction

Implantation of a pacemaker (PM)/cardioverter-defibrillator (ICD), as well as reimplantation and repair operations, is associated with a low risk of complications, which usually occur early after the procedure. Possible complications include the following: PM pocket haematoma, stimulation of the chest wall muscle, diaphragm, phrenic nerves, lead dislocations within the heart, infections (bacterial endocarditis of the tricuspid valve, of the lead tip area, pacemaker pocket), subclavian vein thrombosis and/or its supplies, and in rare cases right ventricle or right atrium perforation with possible tamponade symptoms. The most common late complication is lead fracture, which requires its repair or removal. Another possible complication is PM dislocation under the skin (which may cause decubitus or even skin necrosis) and skin puncture by a lead or PM [1]. Heart perforation in patients after PM implantation is a rare but potentially fatal complication. It may involve large veins, atrial or ventricular walls or the coronary sinus. Perforations may be acute (up to 24 h after implantation, which may lead to tamponade or death), subacute, which occur up to 1 month after implantation, or chronic, which appear at a later time. Another classification distinguishes between early (symptoms occur up to 1 month after implantation) and delayed perforations. The estimated rate of this complication is 1% and depends mainly on the surgeon's experience [2-13]. Main risk factors are listed in Table I [1, 4, 5].
Table I

Risk factors of lead perforation

 I. Type of lead, mode of stimulation, lead tip location
  Temporary stimulation
  Atrial stimulation
  Active fixation
  Defibrillator leads:

Double spirals

Number of shocks delivered

Excessive length of the lead

Small diameter of the lead

High resistance (small tip surface)

Apical position

 II. Propriety of the heart muscle
  Congestive cardiomyopathy
 III. The elderly, women, patients with low body mass (< 20 kg/m2)
 IV. Anticoagulation therapy and steroid use within 7 days of implantation
 V. Chest trauma, especially soon after implantation
Risk factors of lead perforation Double spirals Number of shocks delivered Excessive length of the lead Small diameter of the lead High resistance (small tip surface) Apical position The paper presents cases of life-threatening subacute heart tamponade and delayed perforations in patients after PM/ICD and atrial/ventricular lead implantation.

Material and methods

In our clinic in 2005–2010, 6 patients with heart perforation were hospitalized (3 women, 3 men), mean age 58.6 ±20.8 years (17 to 73 years). In 2 of the patients, implantation was performed outside our clinic. During this period 3600 PM and 1007 ICD implantations were performed. Characteristics of the patient population are presented in Table II.
Table II

Characteristics of the patient population

ParameterPatient 1Patient 2Patient 3Patient 4Patient 5Patient 6
Age [years]707367606517
SexMaleMaleFemaleFemaleMaleFemale
Tachy-brady syndrome+++
Second-degree atrioventricular block, advanced with losses of consciousness+
Vaso-vagal syndrome type IIB with asystole lasting 12 s+
Recurrent non-sustained ventricular tachycardia+
Dual chamber pacemaker+++++
Dual chamber ICD+
Coronary heart disease+++++
Arterial hypertension+++++
Dyslipidaemia+++
Hyperthyreosis++
Diabetes mellitus++
Obesity+
Characteristics of the patient population We analysed patients’ medical records, physical examination, X-rays, echocardiography, computed tomography (CT) and protocols of the implantation procedures.

Results

In the present study, the incidence of heart perforation was 0.09%. Symptoms developed 4 to 990 days (mean 186.3 ±394.3) after device implantation. There was no case of acute heart tamponade/perforation up to 24 h after implantation. The perforation site was found in the right atrial wall in one case and the right ventricular wall in 6 cases (Figures 1–2). Ventricular leads were placed in the right ventricular outflow tract and in 1 case in the apex. Atrial leads were placed in the right auricle. Heart wall damage was caused by pacing ventricular and atrial leads (Biotronik Setrox S60, S53, Vitatron Crystalline, Medtronic CapSureFixNovus) in 5 cases and by a 1-coil defibrillator lead (Medtronic Sprint Quattro) in one case. All leads were active fixation. Patients complained of stabbing pain in the left side of the chest, increased by movement (in 5 cases), dyspnoea (1 case), diaphragm stimulation (1 case), or remained asymptomatic (1 case). Standard follow-up of the electrical parameters showed pacing failure (in 4 patients), and sensing failure (in 6 cases). Transthoracic echocardiography showed pericardial effusion over 10 mm behind the posterior wall of the left ventricle in the parasternal long-axis view in all patients (Figure 3). Computed tomography scanning confirmed heart perforation (of the atrium and in 6 cases of the ventricle). In the operating room, with full cardiac surgery and anaesthesiology backup, the whole device was removed by direct traction, with pericardiocentesis when necessary (pericardium drainage in 3 cases) in 4 patients, and in one case (Figure 4) cardiac surgical extraction was used (in this patient, the lead penetrated the pericardial sac, injuring the parietal pleura and ended in the intercostal muscles, but without pneumothorax); 1 patient with delayed perforation underwent device removal by percutaneous lead extraction outside our department in a centre specializing in percutaneous lead extractions. Five patients had a new PM/ICD implanted, with lead tip location different from the previous one. One patient (patient no. 6) did not agree to have a new PM implanted.
Figure 1

Transthoracic echocardiography view. Ventricular lead penetrating the wall of the right ventricle

Figure 2

Computed tomography scan. Tips of both atrial and ventricular leads outside the heart

Figure 3

Transthoracic echocardiography view. Pericardial effusion up to 29 mm behind the left ventricular lateral wall

Figure 4

Computed tomography scan. Ventricular lead perforating pericardial sac, parietal pleura and ending in intercostal muscles

Transthoracic echocardiography view. Ventricular lead penetrating the wall of the right ventricle Computed tomography scan. Tips of both atrial and ventricular leads outside the heart Transthoracic echocardiography view. Pericardial effusion up to 29 mm behind the left ventricular lateral wall Computed tomography scan. Ventricular lead perforating pericardial sac, parietal pleura and ending in intercostal muscles

Discussion

Heart perforation is a rare, but potentially fatal complication of permanent cardiac pacing systems. Subacute and chronic perforation may be asymptomatic, or it may cause dangerous symptoms resulting in death [2, 3, 9, 11, 13]. The common symptoms are chest pain, chest muscle or diaphragm contraction, hiccough, abdominal pain, dyspnoea, faintness or losses of consciousness (because of pacing failure), inadequate shocks of the ICD, symptoms of heart failure or tamponade [2, 3, 9, 12, 13]. Heart tamponade is a life-threatening complication, caused by rapid blood accumulation in the pericardial sac. It leads to pressure growth in the pericardium, which in turn impairs blood return to the atria and ventricles, lowering ejection volume and causing haemodynamic collapse with symptoms of cardiogenic shock, arrhythmias or death [10]. In our population, there was no case of an acute heart tamponade. The most common perforation symptoms described in the literature are improper electrophysiological parameters; however, there have been cases with normal electrical function [1]. Failure of the lead to pace (increased pacing threshold) or sense (decreased R wave amplitude) or decreased lead impedance may suggest perforation and require further diagnosis [5, 7]. In our material, in each case we observed a high pacing or sensing threshold. Diagnostic methods in heart perforation involve chest X-ray and fluoroscopy (to show lead dislocation outside the border of the heart), transthoracic and transoesophageal echocardiography (TTE and TEE) (presence of the lead tip in the pericardium or pericardial effusion). The most important diagnostic method remains CT, which is the gold standard. It reveals both lead dislocation and even small pericardial effusion [4, 9, 12, 13]. Each of our patients had pericardial effusion. In case of symptomatic heart perforation, the best method of treatment is extraction of the whole device. Various techniques of lead removal are used depending on the lead construction and the time since its implantation [5]. Active fixation leads of homogeneous diameter up to 2 years after implantation (6 months in the case of defibrillator leads) may be removed by direct traction, as in 4 of our patients. In case of failure, Byrd dilatators are used. Active fixation leads of uneven diameter and/or implanted over 2 years before or other lead types (passive fixation) and implanted over 1 year before (6 months in the case of defibrillator leads) should be removed by a classical set of Byrd dilatators [1, 5, 14], as occurred in our 1 patient with delayed perforation. Most authors believe that patients with heart perforation should be transported to a centre specializing in percutaneous lead extraction and the best treatment is direct traction or percutaneous lead extraction in the operating room, with monitoring of basic life functions. Cardiosurgical, anaesthesiological and echocardiographic teams ready at hand are always necessary [1, 5]. It is also advised to have TEE monitoring during and after the operation. Percutaneous extraction of the perforating lead, not requiring general anaesthesia, has been performed in cases of a very high operation risk. American guidelines recommend surgical removal of the perforating leads [6, 8]. The reported cases of surgical lead removal were digestive tract perforation, massive heart tamponade, and atypical location of the perforating lead in the intercostal muscles [9, 12], as in our patient who required cardiac surgical operation. In most of these cases, a new pacemaker was implanted, placing the lead tip in a different location than before. In the case of open surgery, epicardial leads have been used [4]. Diagnosis should be based on the patient's medical records, telemetric control of electrical parameters and visual examinations. Computed tomography remains the gold standard and it confirmed perforation in all our patients. The main limitation of this study is the small group: only 6 patients in 5 years. None of the above-mentioned predictors appeared in our patients with a significant frequency. In conclusion, in case of symptomatic heart perforation, the best method of treatment is extraction of the whole device. The perforating lead may be removed by direct traction in the operating room with cardiosurgical, anaesthesiological and echocardiographic backup. In case of lead perforation outside the pericardial sac or its atypical location, cardiac surgery is a safer method. The most important diagnostic method remains CT.
  11 in total

1.  [PM/ICD lead extraction - most difficult and potentially hazardous electrotherapy procedure - logistic and training problems].

Authors:  Andrzej Kutarski; Grzegorz Opolski
Journal:  Kardiol Pol       Date:  2010-06       Impact factor: 3.108

Review 2.  Delayed perforation of the right ventricle as a complication of permanent cardiac pacing - is following the guidelines always the right choice? Non-standard treatment - a case report and literature review.

Authors:  Anna Rydlewska; Barbara Małecka; Andrzej Zabek; Piotr Klimeczek; Jacek Lelakowski; Mieczysław Pasowicz; Marek Czajkowski; Andrzej Kutarski
Journal:  Kardiol Pol       Date:  2010-03       Impact factor: 3.108

Review 3.  Delayed lead perforation: a disturbing trend.

Authors:  Mohammed N Khan; George Joseph; Yaariv Khaykin; Khaled M Ziada; Bruce L Wilkoff
Journal:  Pacing Clin Electrophysiol       Date:  2005-03       Impact factor: 1.976

4.  Uncomplicated heart and lung perforation by a displaced ventricular pacemaker lead: a case report.

Authors:  Hatice Selcuk; M Timur Selcuk; Orhan Maden; Ozcan Ozeke; M Kutlu Celenk; Aysel Turkvatan; Sule Korkmaz
Journal:  Pacing Clin Electrophysiol       Date:  2006-04       Impact factor: 1.976

5.  Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA).

Authors:  Bruce L Wilkoff; Charles J Love; Charles L Byrd; Maria Grazia Bongiorni; Roger G Carrillo; George H Crossley; Laurence M Epstein; Richard A Friedman; Charles E H Kennergren; Przemyslaw Mitkowski; Raymond H M Schaerf; Oussama M Wazni
Journal:  Heart Rhythm       Date:  2009-05-22       Impact factor: 6.343

6.  [An old ventricular lead migration and subsequent dislodgement to the pleural cavity - an unusual complication of permanent cardiac pacing].

Authors:  Elzbieta Adamowicz-Czoch; Jan Głowacki; Michał Hawranek; Bogusława Dziubek; Marek Gierlotka; Marian Zembala
Journal:  Kardiol Pol       Date:  2008-08       Impact factor: 3.108

7.  Late presentation of a lead perforation as a complication of permanent pacemaker insertion.

Authors:  Salman A Haq; John F Heitner; Leonard Lee; John T Kassotis
Journal:  Angiology       Date:  2008-04-02       Impact factor: 3.619

8.  [Indications for the procedure for transvenous removing of electrodes based on the guidelines of U.S. societies].

Authors:  Jacek Lelakowski
Journal:  Pol Merkur Lekarski       Date:  2010-03

9.  Management of subacute and delayed right ventricular perforation with a pacing or an implantable cardioverter-defibrillator lead.

Authors:  Julien Laborderie; Laurent Barandon; Sylvain Ploux; Antoine Deplagne; Bilel Mokrani; Sylvain Reuter; François Le Gal; Pierre Jais; Michel Haissaguerre; Jacques Clementy; Pierre Bordachar
Journal:  Am J Cardiol       Date:  2008-09-12       Impact factor: 2.778

10.  Complex percutaneous extraction of a 15-year-old atrial lead dislodged into the subclavian vein.

Authors:  Jacek Lelakowski; Andrzej Kutarski; Barbara Małecka; Jacek Majewski
Journal:  Arch Med Sci       Date:  2011-03-08       Impact factor: 3.318

View more
  10 in total

1.  Right coronary artery perforation by an active-fixation atrial pacing lead resulting in life-threatening tamponade.

Authors:  Eiichiro Nakagawa; Yukio Abe; Ryushi Komatsu; Takahiko Naruko; Akira Itoh
Journal:  J Arrhythm       Date:  2015-03-16

2.  Right heart perforation by pacemaker leads.

Authors:  Marek Banaszewski; Janina Stępińska
Journal:  Arch Med Sci       Date:  2012-02-29       Impact factor: 3.318

3.  Subacute right ventricle perforation by pacemaker lead presenting with left hemothorax and shock.

Authors:  Julianne Nichols; Natalie Berger; Praveen Joseph; Debapriya Datta
Journal:  Case Rep Cardiol       Date:  2015-02-18

4.  Impact of Pacemaker Lead Characteristics on Pacemaker Related Infection and Heart Perforation: A Nationwide Population-Based Cohort Study.

Authors:  Yu-Sheng Lin; Tien-Hsing Chen; Sheng-Ping Hung; Dong Yi Chen; Chun-Tai Mao; Ming-Lung Tsai; Shih-Tai Chang; Chun-Chieh Wang; Ming-Shien Wen; Mien-Cheng Chen
Journal:  PLoS One       Date:  2015-06-15       Impact factor: 3.240

5.  Risk factors influencing complications of cardiac implantable electronic device implantation: infection, pneumothorax and heart perforation: a nationwide population-based cohort study.

Authors:  Yu-Sheng Lin; Sheng-Ping Hung; Pei-Rung Chen; Chia-Hung Yang; Hung-Ta Wo; Po-Cheng Chang; Chun-Chieh Wang; Chung-Chuan Chou; Ming-Shien Wen; Chang-Ming Chung; Tien-Hsing Chen
Journal:  Medicine (Baltimore)       Date:  2014-12       Impact factor: 1.889

6.  Ventricular perforation by pacemaker lead repaired with two hemostatic devices.

Authors:  Filippo Prestipino; Antonio Nenna; Adele Casacalenda; Massimo Chello
Journal:  Int J Surg Case Rep       Date:  2014-10-17

7.  Reversibility of High-Grade Atrioventricular Block with Revascularization in Coronary Artery Disease without Infarction: A Literature Review.

Authors:  Rhanderson Cardoso; Carlos E Alfonso; James O Coffey
Journal:  Case Rep Cardiol       Date:  2016-01-26

Review 8.  Identification and management of right ventricular perforation using pacemaker and cardioverter-defibrillator leads: A case series and mini review.

Authors:  Mohammad Ali Akbarzadeh; Reza Mollazadeh; Salma Sefidbakht; Soraya Shahrzad; Negar Bahrololoumi Bafruee
Journal:  J Arrhythm       Date:  2016-06-30

9.  Delayed right ventricular lead perforation complicated by tamponade in biventricular hypertrophic cardiomyopathy.

Authors:  Najla Kourireche; Amal Boutakhrit; Fatima Chikhi; Ibtissam Fellat; Mohammed Cherti
Journal:  Clin Case Rep       Date:  2017-10-16

10.  Outcomes of transvenous lead extraction in patients with lead perforation: A single-center experience.

Authors:  Xu Zhou; Feng Ze; Ding Li; Long Wang; Jihong Guo; Xuebin Li
Journal:  Clin Cardiol       Date:  2020-01-06       Impact factor: 2.882

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.