[Evaluation of a portable device based on peripheral arterial tone in the detection of obstructive sleep apnea].

Juan Zhang1, Chun-yan Wang, Ning-yu Wang, Jun-qing Liu, Jing-yan Du, Guo-xia Zhou, Zhan-feng Yan, Xiao-hui Wen, Li Li, Jin-feng Liu.   

Abstract

OBJECTIVE: To assess the accuracy of a wrist-worn device (Watch-PAT 200) in the diagnosis of obstructive sleep apnea syndrome (OSAHS).
METHODS: Forty-three adult subjects with suspected OSAHS simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the Watch-PAT 200 during a full-night recording. PSG sleep and respiratory events were scored according to standard criteria. The PSG recordings were blindly manually analyzed, while Watch-PAT data were scored automatically based on the algorithm developed previously.
RESULTS: The mean age of the subjects was (42.2 ± 12.2) years (x(-) ± s), and mean body mass index was (28.0 ± 3.9) kg/m(2). Mean PSG apnea hypopnea index (AHI) was (34.9 ± 29.9) events per hour, and mean PAT-AHI was (36.0 ± 29.2) events per hour. There was a significant correlation between PAT AHI and AHI by PSG (r = 0.931, P < 0.01). A Bland-Altman plot of PAT AHI and PSG AHI was also used to assess the accuracy of Watch-PAT 200. At lower levels of AHI, PAT tended to overestimate disease severity, while at higher levels of AHI, Watch-PAT underestimated severity. To assess sensitivity and specificity of Watch-PAT, constructed receiver operator characteristic curves using a variety of AHI threshold values (5, 15 and 30 events per hour). For AHI ≥ 5 events per hour as threshold value, the Watch-PAT diagnosing rate was 93%, and sensitivity as well as specificity were 94.7% and 80.0%. The misdiagnosis rate and missed diagnosis rate were 20.0% and 5.3%. Optimal combinations of sensitivity and specificity for the AHI threshold values (15 and 30 events per hour) were 82.6% and 100.0%, 95.0% and 95.7% respectively.
CONCLUSION: The Watch-PAT 200 may offer an accurate, robust, and reliable ambulatory method for the detection of OSAHS, with minimal patient discomfort.

Entities:  

Mesh:

Year:  2012        PMID: 22455808

Source DB:  PubMed          Journal:  Zhonghua Er Bi Yan Hou Tou Jing Wai Ke Za Zhi        ISSN: 1673-0860


  1 in total

1.  NExUS-Heart: Novel examinations using smart technologies for heart health-Data sharing from commercial wearable devices and telehealth engagement in participants with or at risk of atrial fibrillation.

Authors:  Sri Nuvvula; Eric Y Ding; Connor Saleeba; Qiming Shi; Ziyue Wang; Alok Kapoor; Jane S Saczynski; Steven A Lubitz; Lara C Kovell; M Diane McKee; David D McManus
Journal:  Cardiovasc Digit Health J       Date:  2021-08-23
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.