Evaluation of the novel bipolar vessel sealing and cutting device BiCision® in a porcine model.

Mara Natascha Szyrach1, Pascal Paschenda, Mamdouh Afify, Daniel Schäller, René Hany Tolba.   

Abstract

BACKGROUND: Energy-based technologies for tissue sealing and cutting are increasingly supplementing current standards used for haemostasis and dissection during laparoscopic surgery. For their safe and efficacious use in clinical practice, these instruments have to guarantee sufficient burst resistance and low thermal damage to adjacent tissue in combination with good cutting characteristics.
MATERIAL AND METHODS: The novel laparoscopic, bipolar electrosurgical sealing and cutting instrument BiCision® was compared to a commercially available laparoscopic device (EnSeal(™)) on visceral and peripheral arteries and veins in an animal model.
RESULTS: For all parameters investigated (burst pressure, cut quality, tissue adhering to the instrument, time needed to seal and cut the vessel and thermal damage), BiCision® was at least as good as EnSeal(™). Regarding the burst pressure, BiCision® was superior over EnSeal(™) in arteries: 600 mmHg (±478) versus 241 (±269) mmHg, respectively (p < 0.0001*). In veins, almost equivalent burst pressures of 155 ± 134 mmHg (BiCision®) and 173 ± 139 mmHg (EnSeal(™)) were obtained.
CONCLUSION: BiCision® appeared to be as good as or even superior to EnSeal(™). Since EnSeal(™) has already been shown to be safe and has been successfully used in clinical practice, BiCision® is assumed to be as efficient and reliable as EnSeal(™) under pre-clinical conditions.

Entities:  

Mesh:

Year:  2012        PMID: 22455599     DOI: 10.3109/13645706.2012.661373

Source DB:  PubMed          Journal:  Minim Invasive Ther Allied Technol        ISSN: 1364-5706            Impact factor:   2.442


  3 in total

1.  Sealing of vessels larger than 7 millimeters using Enseal in porcine aorta.

Authors:  Shahida Bibi; Saeed Alblawi; Vamsi Velchuru; Gopichand Thripuraneni; Francisco Quinteros; Jasna Coralic; Slawomir Marecik; John Park; Leela M Prasad
Journal:  JSLS       Date:  2014 Jul-Sep       Impact factor: 2.172

2.  Safety and effectiveness of a novel generator algorithm for bipolar vessel sealing: a randomised controlled chronic animal study.

Authors:  Bernhard Kraemer; Christos Tsaousidis; Stephan Kruck; Martin Schenk; Marcus Scharpf; Stefan Kommoss; Sara Brucker; Daniela Nuessle; Markus D Enderle; Ulrich Biber
Journal:  BMC Surg       Date:  2019-11-05       Impact factor: 2.102

3.  Laparoscopic liver resection in the semiprone position for tumors in the anterosuperior and posterior segments, using a novel dual-handling technique and bipolar irrigation system.

Authors:  Tetsuo Ikeda; Takao Toshima; Norifumi Harimoto; Youichi Yamashita; Toru Ikegami; Tomoharu Yoshizumi; Yuji Soejima; Ken Shirabe; Yoshihiko Maehara
Journal:  Surg Endosc       Date:  2014-03-13       Impact factor: 4.584

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.