Literature DB >> 22445160

Effect of stem length on prosthetic radial head micromotion.

Dave R Shukla1, James S Fitzsimmons, Kai-Nan An, Shawn W O'Driscoll.   

Abstract

BACKGROUND: Osteointegration of press-fit radial head implants is achieved by limiting micromotion between the stem and bone. Aspects of stem design that contribute to the enhancement of initial stability (ie, stem diameter and surface coating) have been investigated. The importance of total prosthesis length and level of the neck cut has not been examined.
METHODS: Cadaveric radii were implanted with cementless, porous-coated radial head stems. We resected 10, 12, 15, 20, and 25 mm of radial neck in each specimen. Stem-bone micromotion was measured after each cut. Values were expressed in terms of quotients (cantilever quotient).
RESULTS: A threshold effect was observed at 15 mm of neck resection (cantilever quotient, 0.4), with a significant increase in micromotion observed between 12 mm (40 ± 10 μm) and 15 mm (80 ± 25 μm). A cantilever quotient of 0.35 or less predicted implant stability, whereas implants with a cantilever quotient of 0.6 or more were unstable. In between, the stems were "at risk" of instability.
CONCLUSION: Initial stem stability of a porous-coated, cementless radial head implant is dependent on length of the implant stem within bone and the level of the cut (amount of bone resected). Stability may be compromised by an implant with a combined head and neck length that is too long compared with the stem length within the canal. We found a critical ratio of exposed prosthesis to total implant length (cantilever quotient of 0.4), which puts the prosthesis at risk of inadequate initial stability. These data carry important implications for implant design and use.
Copyright © 2012 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Mosby, Inc. All rights reserved.

Mesh:

Year:  2012        PMID: 22445160     DOI: 10.1016/j.jse.2011.11.025

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  12 in total

1.  Mid-term clinical outcome comparison of long-stemmed monopolar osseointegrated and short-stemmed bipolar radial head prostheses.

Authors:  Yves Gramlich; Eva Krausch; Thomas Stein; Kai Schmidt-Horlohé; Reinhard Hoffmann; Alexander Klug
Journal:  Arch Orthop Trauma Surg       Date:  2020-07-27       Impact factor: 3.067

2.  [Endoprosthetics of acute radial head fractures].

Authors:  K J Burkhart; K Wegmann; C Ries; L P Müller
Journal:  Unfallchirurg       Date:  2013-08       Impact factor: 1.000

3.  Prosthetic radial head stem pull-out as a mode of failure: a biomechanical study.

Authors:  Dave Shukla; James Fitzsimmons; Kai-Nan An; Shawn O'Driscoll
Journal:  Int Orthop       Date:  2013-09-17       Impact factor: 3.075

4.  Causes for early and late surgical re-intervention after radial head arthroplasty.

Authors:  Pierre Laumonerie; David Ancelin; Nicolas Reina; Meagan E Tibbo; Panagiotis Kerezoudis; Stephanie Delclaux; Nicolas Bonnevialle; Pierre Mansat
Journal:  Int Orthop       Date:  2017-05-12       Impact factor: 3.075

5.  Short to midterm outcomes of one hundred and seventy one MoPyC radial head prostheses: meta-analysis.

Authors:  Pierre Laumonerie; Meagan E Tibbo; Panagiotis Kerezoudis; Marc Olivier Gauci; Nicolas Reina; Nicolas Bonnevialle; Pierre Mansat
Journal:  Int Orthop       Date:  2018-07-30       Impact factor: 3.075

Review 6.  Radial head arthroplasty: a historical perspective.

Authors:  Pierre Laumonerie; Meagan E Tibbo; Nicolas Reina; Thuy Trang Pham; Nicolas Bonnevialle; Pierre Mansat
Journal:  Int Orthop       Date:  2018-08-10       Impact factor: 3.075

Review 7.  [Primary radial head arthroplasty in trauma : Complications].

Authors:  K Schmidt-Horlohé; S Buschbeck; D Wincheringer; M Weißenberger; R Hoffmann
Journal:  Orthopade       Date:  2016-10       Impact factor: 1.087

8.  Tight-fitting radial head prosthesis: does stem size help prevent painful loosening?

Authors:  Pierre Laumonerie; Nicolas Reina; Claudia Gutierrez; Stephanie Delclaux; Meagan E Tibbo; Nicolas Bonnevialle; Pierre Mansat
Journal:  Int Orthop       Date:  2017-09-21       Impact factor: 3.075

9.  Monobloc radial head prostheses in complex elbow injuries: results after primary and secondary implantation.

Authors:  Jan Christoph Katthagen; Gunnar Jensen; Helmut Lill; Christine Voigt
Journal:  Int Orthop       Date:  2012-12-28       Impact factor: 3.075

10.  Radial head replacement with unipolar and bipolar SBi system: a clinical and radiographic analysis after a 2-year mean follow-up.

Authors:  Roberto Rotini; Alessandro Marinelli; Enrico Guerra; Graziano Bettelli; Michele Cavaciocchi
Journal:  Musculoskelet Surg       Date:  2012-04-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.