| Literature DB >> 22444430 |
Hui-Chao Hu1, Hui-Jun Jin, Xin-Sheng Chai.
Abstract
A headspace gas chromatographic (HS-GC) method has been developed for the determination of residual hydrogen peroxide in pulp bleaching effluents. The method is based on the reaction of hydrogen peroxide and permanganate in an acidic medium (0.1 mol/L), in which hydrogen peroxide is quantitatively converted to oxygen within 10 min at 60°C in a sealed headspace sample vial. The released oxygen is then determined by GC equipped with a thermal conductivity detector. The method is robust, sensitive, and accurate, with reproducibility characterized by a relative standard deviation of <0.5%, a sensitivity whose limit of quantification (LOQ) is 0.96 μmol, and a demonstrated recovery ranging from 98 to 103%. Further, the method is simple, rapid, and automated.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22444430 DOI: 10.1016/j.chroma.2012.02.069
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759