Effects of atmospheric pressure conditions on flow rate of an elastomeric infusion pump.

Jong Wang1, Anna Moeller, Yuanpang Samuel Ding.   

Abstract

PURPOSE: The effects of pressure conditions, both hyperbaric and hypobaric, on the flow rate of an elastomeric infusion pump were investigated.
METHODS: The altered pressure conditions were tested with the restrictor outlet at two different conditions: (1) at the same pressure condition as the Infusor elastomeric balloon and (2) with the outlet exposed to ambient conditions. Five different pressure conditions were tested. These included ambient pressure (98-101 kilopascals [kPa]) and test pressures controlled to be 10 or 20 kPa below or 75 or 150 kPa above the ambient pressure. A theoretical calculation based on the principles of fluid mechanics was also used to predict the pump's flow rate at various ambient conditions.
RESULTS: The conditions in which the Infusor elastomeric pump and restrictor outlet were at the same pressure gave rise to average flow rates within the ±10% tolerance of the calculated target flow rate of 11 mL/hr. The flow rate of the Infusor pump decreased when the pressure conditions changed from hypobaric to ambient. The flow rate increased when the pressure conditions changed from hyperbaric to ambient.
CONCLUSION: The flow rate of the Infusor elastomeric pump was not affected when the balloon reservoir and restrictor outlet were at the same pressure. The flow rate varied from 58.54% to 377.04% of the labeled flow rate when the pressure applied to the reservoir varied from 20 kPa below to 150 kPa above the pressure applied to the restrictor outlet, respectively. The maximum difference between observed flow rates and those calculated by applying fluid mechanics was 4.9%.

Mesh:

Substances:

Year:  2012        PMID: 22441790     DOI: 10.2146/ajhp110296

Source DB:  PubMed          Journal:  Am J Health Syst Pharm        ISSN: 1079-2082            Impact factor:   2.637


  2 in total

1.  An assessment of the performance of the Baxter elastomeric (LV10) Infusor™ pump under hyperbaric conditions.

Authors:  Stephen Perks; Denise F Blake; Derelle A Young; John Hardman; Lawrence H Brown; Iestyn Lewis; Tilley Pain
Journal:  Diving Hyperb Med       Date:  2017-03       Impact factor: 0.887

Review 2.  Ambulatory chemotherapy: Past, present, and future.

Authors:  Racha Sabbagh Dit Hawasli; Stephen Barton; Shereen Nabhani-Gebara
Journal:  J Oncol Pharm Pract       Date:  2021-01-18       Impact factor: 1.809

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.