Hyperthermia in bone generated with MR imaging-controlled focused ultrasound: control strategies and drug delivery.

Robert Staruch1, Rajiv Chopra, Kullervo Hynynen.   

Abstract

PURPOSE: To evaluate the feasibility of achieving image-guided drug delivery in bone by using magnetic resonance (MR) imaging-controlled focused ultrasound hyperthermia and temperature-sensitive liposomes.
MATERIALS AND METHODS: Experiments were approved by the institutional animal care committee. Hyperthermia (43°C, 20 minutes) was generated in 10-mm-diameter regions at a muscle-bone interface in nine rabbit thighs by using focused ultrasound under closed-loop temperature control with MR thermometry. Thermosensitive liposomal doxorubicin was administered systemically during heating. Heating uniformity and drug delivery were evaluated for control strategies with the temperature control image centered 10 mm (four rabbits) or 0 mm (five rabbits) from the bone. Simulations estimated temperature elevations in bone. Drug delivery was quantified by using the fluorescence of doxorubicin extracted from bone marrow and muscle and was compared between treated and untreated thighs by using the one-sided Wilcoxon signed rank test.
RESULTS: With ultrasound focus and MR temperature control plane 0 mm and 10 mm from the bone interface, average target region temperatures were 43.1°C and 43.3°C, respectively; numerically estimated bone temperatures were 46.8°C and 78.1°C. The 10-mm offset resulted in thermal ablation; numerically estimated muscle temperature was 66.1°C at the bone interface. Significant increases in doxorubicin concentration occurred in heated versus unheated marrow (8.2-fold, P = .002) and muscle (16.8-fold, P = .002). Enhancement occurred for 0- and 10-mm offsets, which suggests localized drug delivery in bone is possible with both hyperthermia and thermal ablation.
CONCLUSION: MR imaging-controlled focused ultrasound can achieve localized hyperthermia in bone for image-guided drug delivery in bone with temperature-sensitive drug carriers. © RSNA, 2012.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22438444     DOI: 10.1148/radiol.11111189

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  16 in total

Review 1.  MRI-guided focused ultrasound surgery in musculoskeletal diseases: the hot topics.

Authors:  Alberto Bazzocchi; Alessandro Napoli; Beatrice Sacconi; Giuseppe Battista; Giuseppe Guglielmi; Carlo Catalano; Ugo Albisinni
Journal:  Br J Radiol       Date:  2015-11-26       Impact factor: 3.039

2.  Experimental assessment of microwave ablation computational modeling with MR thermometry.

Authors:  Pegah Faridi; Paul Keselman; Hojjatollah Fallahi; Punit Prakash
Journal:  Med Phys       Date:  2020-07-16       Impact factor: 4.071

3.  Targeted antibiotic delivery using low temperature-sensitive liposomes and magnetic resonance-guided high-intensity focused ultrasound hyperthermia.

Authors:  Rachel Wardlow; Chenchen Bing; Joshua VanOsdol; Danny Maples; Michelle Ladouceur-Wodzak; Michele Harbeson; Joris Nofiele; Robert Staruch; Akhilesh Ramachandran; Jerry Malayer; Rajiv Chopra; Ashish Ranjan
Journal:  Int J Hyperthermia       Date:  2016-02-18       Impact factor: 3.914

Review 4.  Thermosensitive liposomes for localized delivery and triggered release of chemotherapy.

Authors:  Terence Ta; Tyrone M Porter
Journal:  J Control Release       Date:  2013-04-11       Impact factor: 9.776

5.  Non-invasive measurement of the temperature rise in tissue surrounding a kidney stone subjected to ultrasonic propulsion.

Authors:  Ghanem F Oweis; Barbrina L Dunmire; Bryan W Cunitz; Michael R Bailey
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2015

Review 6.  Lipid-Based Drug Delivery Systems in Cancer Therapy: What Is Available and What Is Yet to Come.

Authors:  Phatsapong Yingchoncharoen; Danuta S Kalinowski; Des R Richardson
Journal:  Pharmacol Rev       Date:  2016-07       Impact factor: 25.468

7.  Model-based feasibility assessment and evaluation of prostate hyperthermia with a commercial MR-guided endorectal HIFU ablation array.

Authors:  Vasant A Salgaonkar; Punit Prakash; Viola Rieke; Eugene Ozhinsky; Juan Plata; John Kurhanewicz; I-C Joe Hsu; Chris J Diederich
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

8.  Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system.

Authors:  Chenchen Bing; Joris Nofiele; Robert Staruch; Michelle Ladouceur-Wodzak; Yonatan Chatzinoff; Ashish Ranjan; Rajiv Chopra
Journal:  Int J Hyperthermia       Date:  2015-11-05       Impact factor: 3.914

9.  The effect of injected dose on localized tumor accumulation and cardiac uptake of doxorubicin in a Vx2 rabbit tumor model using MR-HIFU mild hyperthermia and thermosensitive liposomes.

Authors:  Bingbing Cheng; Chenchen Bing; Robert M Staruch; Sumbul Shaikh; Michelle Wodzak Staruch; Debra Szczepanski; Noelle S Williams; Theodore W Laetsch; Rajiv Chopra
Journal:  Int J Hyperthermia       Date:  2020       Impact factor: 3.914

Review 10.  MRI-guided monitoring of thermal dose and targeted drug delivery for cancer therapy.

Authors:  Ruchika Fernando; Jon Downs; Danny Maples; Ashish Ranjan
Journal:  Pharm Res       Date:  2013-06-19       Impact factor: 4.200

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.