| Literature DB >> 22437365 |
Parla Astarci1, David Glineur, Joelle Kefer, Jean Renkin, Jean-Louis Vanoverschelde, Gebrine El Khoury.
Abstract
Transapical aortic valve implantation requires puncture of the left ventricle apex and insertion of a 32-French delivery sheath. A critical step in the procedure consists of secure closure of the ventricular apex. We describe 2 cases of apical rupture of 42 transapical aortic valve implantations. Furthermore, we describe the use of a newly designed single circular Teflon pledget that can help to avoid this complication. This pledget provides a more secure and uniform shrinkage of the entire apex to close the defect left by the delivery sheath.Year: 2010 PMID: 22437365 DOI: 10.1097/IMI.0b013e3181cbb364
Source DB: PubMed Journal: Innovations (Phila) ISSN: 1556-9845