Literature DB >> 22435281

Ultrasound confirmation of laryngeal mask airway placement correlates with fiberoptic laryngoscope findings.

Deepak Gupta1, Arvind Srirajakalidindi, Nader Habli, Halim Haber.   

Abstract

BACKGROUND: Correct placement of a laryngeal mask airway (LMA) requires confirmation to appreciate the adequacy of laryngeal seal and pulmonary ventilation.
OBJECTIVES: The present study was designed to assess the feasibility of ultrasound use for confirmation of correct placement of LMA and its correlation with fiberoptic laryngoscopy as a confirmation tool for LMA position.
MATERIALS AND METHODS: 31 ASA I and II patients scheduled for same day surgery under general anesthesia underwent standard general anesthetic technique with AuraOnce or AuraFlex Disposable Laryngeal Mask Airways. The position of the LMA cuff was confirmed by transverse neck ultrasound (USG), and reconfirmed with intra-LMA fiberoptic laryngoscopy (FOL).
RESULTS: The ultrasound grade of LMA position strongly correlated positively with the fiberoptic grade of LMA position (r = 0.92; p < 0.0001). This correlation was obtained immediately after LMA placement, as well as just before LMA removal. The Bland-Altman scatter plot showed insignificant differences between the two grading systems with small and good limits of agreement (-0.63 to +0.57).
CONCLUSION: Ultrasound examination can replace fiberoptic examination for confirmation of the correct placement of an LMA. Additionally, non-invasive ultrasound examination can further give insight into the cause of airway/ventilation events that may be interfering with the LMA placement and ventilation.

Entities:  

Mesh:

Year:  2011        PMID: 22435281

Source DB:  PubMed          Journal:  Middle East J Anaesthesiol        ISSN: 0544-0440


  6 in total

1.  Ultrasound: A promising tool for contemporary airway management.

Authors:  Rakesh Garg; Anju Gupta
Journal:  World J Clin Cases       Date:  2015-11-16       Impact factor: 1.337

2.  Assessment of vocal cord movement by ultrasound in the ICU.

Authors:  Zhengshang Ruan; Rongrong Ren; Wenwen Dong; Junjie Ma; Zhenyu Xu; Yanfei Mao; Lai Jiang
Journal:  Intensive Care Med       Date:  2018-11-20       Impact factor: 17.440

3.  Comparison of leakage test and ultrasound imaging to validate ProSeal supraglottic airway device placement.

Authors:  Sachin E Ajithan; Archana Puri; Mukul C Kapoor
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2020-06-15

4.  To compare clinical versus ultrasound assessment of correct placement of ProSeal Laryngeal mask airway (PLMA): a prospective randomized study.

Authors:  Kanika Rustagi; Rakesh Garg; Sachidanand Jee Bharti; Vinod Kumar; Nishkarsh Gupta; Seema Mishra; Sushma Bhatnagar
Journal:  J Clin Monit Comput       Date:  2021-03-16       Impact factor: 2.502

5.  The place of ultrasonography in confirming the position of the laryngeal mask airway in pediatric patients: an observational study.

Authors:  Sule Arican; Sevgi Pekcan; Gulcin Hacibeyoglu; Merve Yusifov; Sait Yuce; Sema Tuncer Uzun
Journal:  Braz J Anesthesiol       Date:  2021-02-03

6.  Ultrasonography for clinical decision-making and intervention in airway management: from the mouth to the lungs and pleurae.

Authors:  Michael S Kristensen; Wendy H Teoh; Ole Graumann; Christian B Laursen
Journal:  Insights Imaging       Date:  2014-02-12
  6 in total