Literature DB >> 22429859

Superiority of gold versus platinum irrigated tip catheter ablation of the pulmonary veins and the cavotricuspid isthmus: a randomized study comparing tip temperatures and cooling flow requirements.

Markus Linhart1, Ilja Liberman, Jan Wilko Schrickel, Erica Lilian Mittmann-Braun, René Andrié, Florian Stöckigt, Jens Kreuz, Georg Nickenig, Lars Martin Lickfett.   

Abstract

INTRODUCTION: In order to optimize power delivery into the myocardium during radiofrequency ablation (RFA) without overheating the electrode tip, active cooling of the tip electrode as well as electrode tips made of gold have evolved. Recently, an externally irrigated gold tip electrode ablation catheter has been developed to combine the advantages of these 2 technologies. We sought to investigate the procedural parameters tip temperature, delivered power and cooling flow requirements of the irrigated gold tip catheter in comparison to the conventional irrigated platinum iridium (Pt) tip catheter in pulmonary vein isolation (PVI) and cavotricuspid isthmus (CTI) ablation. METHODS AND
RESULTS: Sixty patients referred for first PVI were randomized into ablation with irrigated gold tip catheter versus irrigated Pt tip catheter. Forty-nine patients received ablation of CTI following PVI. Mean and standard deviation from all measurements were calculated for each patient. During RFA of pulmonary veins, mean catheter tip temperature was significantly lower in the gold group (35.4 ± 0.9 °C vs 38.2 ± 0.8 °C, P < 0.001), and total amount of delivered energy was higher (1303.1 ± 81.1 W vs 1223.7 ± 115.6 W, P = 0.004). During CTI ablation, necessary saline flow was almost 2.5-fold lower in the gold group (22.5 ± 5.9 mL/min vs 52.5 ± 9.7 mL/min, P < 0.001), accompanied by significantly lower tip temperature (39.1 ± 0.6 °C vs 40.5 ± 1.4 °C, P < 0.001).
CONCLUSION: The irrigated gold tip electrode allows to deliver significantly more energy at a lower electrode tip temperature in RFA of PV and CTI in comparison to the irrigated Pt tip electrode. The required saline flow during CTI ablation is much lower than in Pt.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22429859     DOI: 10.1111/j.1540-8167.2011.02267.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  3 in total

Review 1.  Global Substrate Mapping and Targeted Ablation with Novel Gold-tip Catheter in De Novo Persistent AF.

Authors:  Michael Tb Pope; Timothy R Betts
Journal:  Arrhythm Electrophysiol Rev       Date:  2022-04

2.  Silent cerebral lesions and cognitive function after pulmonary vein isolation with an irrigated gold-tip catheter: REDUCE-TE Pilot study.

Authors:  Boris Schmidt; Gábor Széplaki; Bela Merkely; Josef Kautzner; Vincent van Driel; Felix Bourier; Malte Kuniss; Alan Bulava; Georg Nölker; Muchtiar Khan; Thorsten Lewalter; Norbert Klein; Beate Wenzel; Julian Kr Chun; Dipen Shah
Journal:  J Cardiovasc Electrophysiol       Date:  2019-03-25

3.  First clinical experience of high-power ablation of atrial fibrillation with a novel contact force-sensing gold-tip catheter.

Authors:  Abdul Shokor Parwani; Adrian Jayanata; Robin Kraft; Philipp Lacour; Florian Blaschke; Burkert Pieske; Leift-Hendrik Boldt
Journal:  Cardiol J       Date:  2022-06-28       Impact factor: 3.487

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.