Accurate activity measurement of Lu-177 by the liquid scintillation 4πβ-γ coincidence counting technique.

B R S Simpson1, M J van Staden, J Lubbe, W M van Wyngaardt.   

Abstract

This paper reports on absolute measurements made at the NMISA by the 4π[LS]β-γ coincidence extrapolation technique, which formed part of a key comparison of activity measurements of the radionuclide lutetium-177, a beta-gamma emitter. A detection efficiency analysis based on the decay scheme indicated that the coincidence extrapolation method is feasible for extracting the source disintegration rate of (177)Lu. A simulation was undertaken utilizing the efficiency equations to give an indication of the expected variation of the source count rate with efficiency for different gamma-ray windows. Both the simulation and experiment indicated that the window giving the highest counting efficiency was centered about the 208.4 keV photopeak, with the data showing linear variation in the upper efficiency range. The extracted activity concentration of the (177)Lu solution proved to be highly accurate, being higher than the proposed comparison reference value by 0.15%, well within one standard uncertainty (σ=0.26%) as specified by the NMISA.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22424836     DOI: 10.1016/j.apradiso.2012.02.102

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Monte Carlo based approach to the LS–NaI 4πβ–γ anticoincidence extrapolation and uncertainty.

Authors:  R Fitzgerald
Journal:  Appl Radiat Isot       Date:  2016-03       Impact factor: 1.513

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.