Literature DB >> 22421418

Optimization of x-ray spectra in digital mammography through Monte Carlo simulations.

D M Cunha1, A Tomal, M E Poletti.   

Abstract

In this work, a Monte Carlo code was used to investigate the performance of different x-ray spectra in digital mammography, through a figure of merit (FOM), defined as FOM = CNR²/(¯)D(g), with CNR being the contrast-to-noise ratio in image and [Formula: see text] being the average glandular dose. The FOM was studied for breasts with different thicknesses t (2 cm ≤ t ≤ 8 cm) and glandular contents (25%, 50% and 75% glandularity). The anode/filter combinations evaluated were those traditionally employed in mammography (Mo/Mo, Mo/Rh, Rh/Rh), and a W anode combined with Al or K-edge filters (Zr, Mo, Rh, Pd, Ag, Cd, Sn), for tube potentials between 22 and 34 kVp. Results show that the W anode combined with K-edge filters provides higher values of FOM for all breast thicknesses investigated. Nevertheless, the most suitable filter and tube potential depend on the breast thickness, and for t ≥ 6 cm, they also depend on breast glandularity. Particularly for thick and dense breasts, a W anode combined with K-edge filters can greatly improve the digital technique, with the values of FOM up to 200% greater than that obtained with the anode/filter combinations and tube potentials traditionally employed in mammography. For breasts with t < 4 cm, a general good performance was obtained with the W anode combined with 60 μm of the Mo filter at 24-25 kVp, while 60 μm of the Pd filter provided a general good performance at 24-26 kVp for t = 4 cm, and at 28-30 and 29-31 kVp for t = 6 and 8 cm, respectively.

Entities:  

Mesh:

Year:  2012        PMID: 22421418     DOI: 10.1088/0031-9155/57/7/1919

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Towards standardization of x-ray beam filters in digital mammography and digital breast tomosynthesis: Monte Carlo simulations and analytical modelling.

Authors:  Suman Shrestha; Srinivasan Vedantham; Andrew Karellas
Journal:  Phys Med Biol       Date:  2017-01-11       Impact factor: 3.609

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.