Literature DB >> 22410036

Cardiac output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation.

Y-F Tsai1, B-C Su, C-C Lin, F-C Liu, W-C Lee, H-P Yu.   

Abstract

BACKGROUND: The upgraded third-generation software (version 3.02) for the FloTrac/Vigileo system has been developed to particularly improve the accuracy of cardiac output (CO) measurements in hyperdynamic conditions. The aim of our study was to compare the CO values obtained using the FloTrac/Vigileo system during orthotopic liver transplantation (OLT) with those obtained in the same circumstances using a Swan-Ganz catheter (bolus thermodilution method).
METHODS: Twenty consecutive recipients scheduled for OLT were studied. Simultaneous CO values measured by both devices were obtained at 10 predefined time points throughout the surgery. A percentage error of not more than 30% was established as the criterion for device interchangeability.
RESULTS: A total of 200 paired measurements were obtained. The CO values derived from the FloTrac/Viligeo ranged from 2.8 to 10.9 L/min, with a mean of 5.91±1.81 L/min. The values from bolus thermodilution ranged from 2.2 to 13.2 L/min, with a mean of 6.12±2.07 L/min. The bias was 0.22, and the limits of agreement were -3.13 to 3.56 L/min. The percentage error between the FloTrac/Viligeo and bolus thermodilution measurements was 54.93%. The percentage errors of paired measurements in three surgical phases by subgroup analysis were 43.50% (dissecting phase), 62.9% (anhepatic phase), and 56.05% (reperfusion phase), respectively.
CONCLUSION: CO measurements obtained using the less invasive arterial waveform FloTrac/Vigileo system upgraded with the third-generation software had poor intraoperative agreement with pulmonary artery thermodilution CO measurements in patients undergoing OLT. Copyright Â
© 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22410036     DOI: 10.1016/j.transproceed.2011.12.045

Source DB:  PubMed          Journal:  Transplant Proc        ISSN: 0041-1345            Impact factor:   1.066


  7 in total

1.  Comparison of an advanced minimally invasive cardiac output monitoring with a continuous invasive cardiac output monitoring during lung transplantation.

Authors:  Roland Tomasi; Stephan Prueckner; Stephan Czerner; Renè Schramm; Gerhard Preissler; Bernhard Zwißler; Vera von Dossow-Hanfstingl
Journal:  J Clin Monit Comput       Date:  2015-07-30       Impact factor: 2.502

Review 2.  The relationship of perioperative fluid administration to outcomes in colorectal and pancreatic surgery: a review of the literature.

Authors:  Oliver S Eng; Laleh G Melstrom; Darren R Carpizo
Journal:  J Surg Oncol       Date:  2015-02-02       Impact factor: 3.454

Review 3.  Hemodynamic monitoring during liver transplantation: A state of the art review.

Authors:  Mona Rezai Rudnick; Lorenzo De Marchi; Jeffrey S Plotkin
Journal:  World J Hepatol       Date:  2015-06-08

4.  The comparison between stroke volume variation and filling pressure as an estimate of right ventricular preload in patients undergoing renal transplantation.

Authors:  Daisuke Toyoda; Mitsue Fukuda; Ririko Iwasaki; Takashi Terada; Nobukazu Sato; Ryoichi Ochiai; Yoshifumi Kotake
Journal:  J Anesth       Date:  2014-06-25       Impact factor: 2.078

5.  Agreement between continuous cardiac output measured by the fourth-generation FloTrac/Vigileo system and a pulmonary artery catheter in adult liver transplantation.

Authors:  Yutaka Murata; Takumi Imai; Chikashi Takeda; Toshiyuki Mizota; Shuji Kawamoto
Journal:  Sci Rep       Date:  2022-07-01       Impact factor: 4.996

6.  Intraoperative hemodynamics monitoring in a patient with pheochromocytoma multisystem crisis: a case report.

Authors:  Atsuko Kato; Hisayoshi Tamai; Kanji Uchida; Takamitsu Ikeda; Yoshitsugu Yamada
Journal:  JA Clin Rep       Date:  2018-05-01

7.  Beat-to-Beat Tracking of Pulse Pressure and Its Respiratory Variation Using Heart Sound Signal in Patients Undergoing Liver Transplantation.

Authors:  Yong-Seok Park; Young-Jin Moon; Sung-Hoon Kim; Jae-Man Kim; Jun-Gol Song; Gyu-Sam Hwang
Journal:  J Clin Med       Date:  2019-04-30       Impact factor: 4.241

  7 in total