Analysis of biological tissues in infant chest for the development of an equivalent radiographic phantom.

D R Pina1, Rafael T F Souza, Sergio B Duarte, Matheus Alvarez, Jose R A Miranda.   

Abstract

PURPOSE: The main purpose of the present study was to determine the amounts of different tissues in the chest of the newborn patient (age ≤1 year), with the aim of developing a homogeneous phantom chest equivalent. This type of phantom is indispensable in the development of optimization procedures for radiographic techniques, including dosimetric control, which is a crucial aspect of pediatric radiology. The authors present a systematic set of procedures, including a computational algorithm, to estimate the amounts of tissues and thicknesses of the corresponding simulator material plates used to construct the phantom.
METHODS: The Gaussian fit of computed tomographic (CT) analysis was applied to classify and quantify different biological tissues. The methodology is summarized with a computational algorithm, which was used to quantify tissues through automated CT analysis. The thicknesses of the equivalent homogeneous simulator material plates were determined to construct the phantom.
RESULTS: A total of 180 retrospective CT examinations with anterior-posterior diameter values ranging 8.5-13.0 cm were examined. The amounts of different tissues were evaluated. The results provided elements to construct a phantom to simulate the infant chest in the posterior-anterior or anterior-posterior (PA/AP) view.
CONCLUSIONS: To our knowledge, this report represents the first demonstration of an infant chest phantom dedicated to the radiology of children younger than one year. This phantom is a key element in the development of clinical charts for optimizing radiographic technique in pediatric patients. Optimization procedures for nonstandard patients were reported previously [Pina et al., Phys. Med. Biol. 49, N215-N226 (2004) and Pina et al., Appl. Radiat. Isot. 67, 61-69 (2009)]. The constructed phantom represents a starting point to obtain radiologic protocols for the infant patient.

Entities:  

Mesh:

Year:  2012        PMID: 22380369     DOI: 10.1118/1.3685588

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Quantification of Pulmonary Inflammatory Processes Using Chest Radiography: Tuberculosis as the Motivating Application.

Authors:  Guilherme Giacomini; José R A Miranda; Ana Luiza M Pavan; Sérgio B Duarte; Sérgio M Ribeiro; Paulo C M Pereira; Allan F F Alves; Marcela de Oliveira; Diana R Pina
Journal:  Medicine (Baltimore)       Date:  2015-07       Impact factor: 1.889

2.  Homogeneous Canine Chest Phantom Construction: A Tool for Image Quality Optimization.

Authors:  Ana Luiza Menegatti Pavan; Maria Eugênia Dela Rosa; Guilherme Giacomini; Fernando Antonio Bacchim Neto; Seizo Yamashita; Luiz Carlos Vulcano; Sergio Barbosa Duarte; José Ricardo de Arruda Miranda; Diana Rodrigues de Pina
Journal:  PLoS One       Date:  2016-04-21       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.