A real-time respiration position based passive breath gating equipment for gated radiotherapy: a preclinical evaluation.

Weigang Hu1, Anjie Xu, Guichao Li, Zhen Zhang, Dave Housley, Jinsong Ye.   

Abstract

PURPOSE: To develop a passive gating system incorporating with the real-time position management (RPM) system for the gated radiotherapy.
METHODS: Passive breath gating (PBG) equipment, which consists of a breath-hold valve, a controller mechanism, a mouthpiece kit, and a supporting frame, was designed. A commercial real-time positioning management system was implemented to synchronize the target motion and radiation delivery on a linear accelerator with the patient's breathing cycle. The respiratory related target motion was investigated by using the RPM system for correlating the external markers with the internal target motion while using PBG for passively blocking patient's breathing. Six patients were enrolled in the preclinical feasibility and efficiency study of the PBG system.
RESULTS: PBG equipment was designed and fabricated. The PBG can be manually triggered or released to block or unblock patient's breathing. A clinical workflow was outlined to integrate the PBG with the RPM system. After implementing the RPM based PBG system, the breath-hold period can be prolonged to 15-25 s and the treatment delivery efficiency for each field can be improved by 200%-400%. The results from the six patients showed that the diaphragm motion caused by respiration was reduced to less than 3 mm and the position of the diaphragm was reproducible for difference gating periods.
CONCLUSIONS: A RPM based PBG system was developed and implemented. With the new gating system, the patient's breath-hold time can be extended and a significant improvement in the treatment delivery efficiency can also be achieved.

Entities:  

Mesh:

Year:  2012        PMID: 22380367     DOI: 10.1118/1.3678986

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Tracking, gating, free-breathing, which technique to use for lung stereotactic treatments? A dosimetric comparison.

Authors:  Jessica Prunaretty; Pierre Boisselier; Norbert Aillères; Olivier Riou; Sebastien Simeon; Ludovic Bedos; David Azria; Pascal Fenoglietto
Journal:  Rep Pract Oncol Radiother       Date:  2018-11-24

2.  Real-time liver tracking algorithm based on LSTM and SVR networks for use in surface-guided radiation therapy.

Authors:  Guangyu Wang; Zhibin Li; Guangjun Li; Guyu Dai; Qing Xiao; Long Bai; Yisong He; Yaxin Liu; Sen Bai
Journal:  Radiat Oncol       Date:  2021-01-14       Impact factor: 3.481

3.  Correlation of Optical Surface Respiratory Motion Signal and Internal Lung and Liver Tumor Motion: A Retrospective Single-Center Observational Study.

Authors:  Guangyu Wang; Xinyu Song; Guangjun Li; Lian Duan; Zhibin Li; Guyu Dai; Long Bai; Qing Xiao; Xiangbin Zhang; Ying Song; Sen Bai
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.